class SetND2(KPFTranslatorFunction):
'''Set the filter in the ND2 filter wheel (the one at the output of the
octagon) via the `kpfcal.ND2POS` keyword.
Args:
CalND2 (str): The neutral density filter to put in the first filter
wheel. This affects only the light injected in to the simultaneous
calibration fiber. Allowed Values: `OD 0.1`, `OD 0.3`, `OD 0.5`,
`OD 0.8`, `OD 1.0`, `OD 4.0`
wait (bool): Wait for move to complete before returning? default: True
KTL Keywords Used:
- `kpfcal.ND2POS`
'''
@classmethod
def pre_condition(cls, args, logger, cfg):
keyword = ktl.cache('kpfcal', 'ND2POS')
allowed_values = list(keyword._getEnumerators())
if 'Unknown' in allowed_values:
allowed_values.pop(allowed_values.index('Unknown'))
check_input(args, 'CalND2', allowed_values=allowed_values)
@classmethod
def perform(cls, args, logger, cfg):
target = args.get('CalND2')
log.debug(f"Setting ND2POS to {target}")
kpfcal = ktl.cache('kpfcal')
kpfcal['ND2POS'].write(target, wait=args.get('wait', True))
@classmethod
def post_condition(cls, args, logger, cfg):
timeout = cfg.getfloat('times', 'nd_move_time', fallback=20)
ND2target = args.get('CalND2')
ND2POS = ktl.cache('kpfcal', 'ND2POS')
if ND2POS.waitFor(f"== '{ND2target}'", timeout=timeout) == False:
raise FailedToReachDestination(ND2POS.read(), ND2target)
@classmethod
def add_cmdline_args(cls, parser, cfg=None):
parser.add_argument('CalND2', type=str,
choices=["OD 0.1", "OD 0.3", "OD 0.5", "OD 0.8",
"OD 1.0", "OD 4.0"],
help='ND2 Filter to use.')
parser.add_argument("--nowait", dest="wait",
default=True, action="store_false",
help="Send move and return immediately?")
return super().add_cmdline_args(parser, cfg)