FCS Issue Requires Additional Mask Design
(2024-Nov-11)
- The MOSFIRE Flexure Compensation System (FCS) saturates more
often than in the past. The saturations appear to coincide with
cold and warm electronics temperatures. The FCS correction is
elevation and rotator angle dependent. As a work around, we
advise observers to create versions of their slitmasks at their
desired PA and also that PA+180 degrees. These "rotated" mask
designs allow FCS to function. In H- and K-bands, please be
aware that the spectral coverage is field-dependent. That is
objects on the left side of the mask have a different wavelength
range than those on the right side in H- and K-bands.
CSU Fatal Errors
(2023-Sep-20)
- MOSFIRE is experiencing more CSU errors recently. Our mitigation
strategies are to move the CSU at EL~67 whenever possible and to
minimize (if possible) CSU moves. One option is to take science
data using the alignment mask which removes the need to move the
bars from alignment to science configuration. This is obviously
an impact on the science, so it is up to the observers to decide
if the tradeoff is worth it. We don't have great statistics on
the fatal errors as they are still a rare occurrence, but in the
limited set of incidents we have seen, doing this at low
elevation (e.g. EL~35) seems to be problematic. Thus if observers are
acquiring a mask at low elevation, we would encourage to either
take data in alignment mask, or to minimize the number of
alignment stars in the mask to minimize the number of bars
moving. If observers chose to carry out observations with
alignment masks, it is worth to design science masks exactly the
sam as alignment masks, that way there is no need to take separate
calibration sets for alignment manually.
From small number statistics that we gathered in August, CSU
errors tend to happen at cos(elevation)*sin(instrument drive
angle)>0.4 and cos(elevation)*cos(instrument drive
angle)<-0.4. These values correspond to large rX and Y
accelerometer values (>6000), where gravity pulls are
higher. Avoiding those combinations of elevation and instrument
drive angle may help reduce the occurrence of CSU errors.
Guider Update (2020-Feb-04)
- The MOSFIRE guider had been experiencing a failure of one of the
amplifiers over the last few months which resulted in the guider field
being reduced to half of the normal field of view. We have replaced the
detector in the guider with a spare which has re-enabled the full field of
view of the guider. This change however, has resulted in some fringing
effects due to the cover glass on the replacement chip. This should have
no effect in most fields, but extremely faint guide stars may be impacted.
We are currently looking to replace the detector with one that does not
suffer from fringing, but this may take some time. A screenshot of the
fringing effect can be seen below.
MAGMA Localization (2019-Jan-24)
- It has been brought to our attention that there is a possible point of
confusion when using MAGMA with certain localization options set on your
computer. MAGMA uses the decimal separator specified by the settings on
the computer (
.
in the US, but which may be ,
for others), but the example input catalog files use the US standard of
the decimal separator. MAGMA users which have European-style localization,
should be aware of this possible point of confusion.
MOSFIRE DRP Version 2018 Released (2018-Jan-05)
- The 2018 release
of the MOSFIRE DRP is out. This new version includes
an easier installation process, python3 compatibility, and an updated
documentation web page, along with numerous other bugfixes and
updates.
Check the online documentation for
installation instructions.
MOSFIRE back on-sky and CSU
operating normally
(2017-Mar-05)
- MOSFIRE is on sky, and taking science data as normal. All
previous restrictions on CSU motions are no longer an
issue. Observers should still be aware, however, avoid
configuring the masks which have physical drive angles of 0 or
180 degrees.
MOSFIRE back on-sky and taking data
(2017-Feb-13)
- MOSFIRE was returned to science operations on 12 Feb
2017. The performance of MOSFIRE so far appears to be the same
as it ever was. Due to weather, we were unable to complete all
of the engineering tests we wanted finish, and plan to complete
those at a later date. We expect that the throughput and
sensitivity of MOSFIRE has not changed.
There is an operational concern with MOSFIRE that observers
should watch when observing. The CSU bar positioning is not as
reliable and we are finding more bars out of position in images
and spectra. Observers should be mindful to check their images
and spectra to ensure bars are okay. Please work with your SA to
help monitor the bar positions and correct the issue in real
time.
For the time being, we have modified the default rotator
angle for acquiring flats to be -10 degrees on the drive angle
(formerly -90). Observer should complete major CSU
configurations at physical drive angles of 10, -10, -170 or -190
to avoid major CSU faults.
MOSFIRE DRP Release
(2016-Dec-12)
- An updated version of the MOSFIRE data reduction package has been
released. This version removes IRAF as a dependency and replaced it with
astropy
and ccdproc
python packages.
More information can be found on the MOSFIRE
DRP web page.
MOSFIRE Repair Mission Status
(2016-Nov-23)
- The MOSFIRE repair is planned for completion by late
January, 2017 and is expected to be ready for science in the new
observing semester beginning February 2017. The MOSFIRE nights
currently on the schedule for January 2017 will need to be
converted to different instruments and possibly different
programs as needed.
The MOSFIRE Dewar was opened in a newly constructed clean room on the
dome floor of Keck I in early November. An optical element in
the collimator assembly was found to be dislodged from its
mount. The optic was shipped back to UCLA for repair and will
be reinstalled by early January. The MOSFIRE Dewar will then be closed
back up, moved up to the Nasmyth deck, pumped and cooled by
mid January where it will undergo daytime testing. An
engineering night will be used in early Feb to confirm on-sky
performance prior to reinstating science operations later in
February.
This plan keeps semester 2017A science programs in tact but we'll be
working with the institutions and individual PI's to change
the remaining MOSFIRE nights in 2016B (January) to other
instruments.
MOSFIRE out of service and
unavailable for science operations
(2016-Sept-15)
- MOSFIRE was removed from science operations due to a
serious problem that is present in the optical system. MOSFIRE will
not be available for science use in October, and we anticipate
the repair mission lasting a few months. MOSFIRE's image quality is too poor to
continue science operations and we are currently scheduling
repair activities. The degradation in image quality was induced
during engineering activities designed to tune MOSFIRE's rotator
bearing. This activity resulted in MOSFIRE vibrating and a loss
in the optical alignment. At the present time, there are many
variables that remain to be constrained to determine where in
the optical path the misalignment occurred, and a deliberate
approach is being taken to complete a repair. Thank you for your
patience during this unfortunate time while we develop our
plan and coordinate a repair mission. Future updates will be
posted to the news page as the repair mission progresses.
Filter wheel one exhibintingexhibiting a problem
(2016-May-25)
- Filter wheel 1 has a repeatable failure mode. When
we move to or from the J2 filter back to the open position, the
filter wheel status goes to unknown. It appears based on the
spectra that the filter wheel is in the correct position, but
the switches are not read correctly to identify the positon for
J2.For most observers, this
should not be an issue as filter wheel 2 is the primary wheel
that is used. Homing will get the wheel out of trouble, but
there are software work arounds we must apply if observers want
to use J2 for science. Please consult with you support
astronomer if you would like to use a medium band filter like
the J2 filter.
Work around for MAGMA saving error
(2016-May-18)
- A couple of possible solutions to the problem described in the previous
news item documenting a MAGMA error are
- Access from the menu bar Tools -> Options and deselect
the first box "Write HTML file when saving slit configuration."
- To run magma offline with
Java version 1.6. A handful of observers had had success with
this workaround. We are working toward updating MAGMA and will
resolve this bug in the next release.
MAGMA experiencing an error on saving
(2016-Apr-26)
- Observers who have recently updated their MAC OS or JAVA are
noticing that MAGMA will not save files correctly. Upon save,
the following error is reported:
"org.xml.sax.SAXParseException;systemId:http://keck.hawaii.edu/mosfire/msc.xsl;
lineNumber: 1; columnNumber: 50; White spaces are required between publicId and systemId."
In addition, the startup script uses xdpyinfo which is an X11
script used to determine the monitor screen size. X11 is no
longer included with the MAC OS as of Jan 2016. These issues do
not appear to prohibit saving *.xml, so
these errors are not preventing mask design and creation.
DVD writer no longer available for data backups
(2016-Jan-26)
- The Keck Observatory is phasing out the auto DVD writer
sometimes used by observers for data backups. Options to
launch the DVD autobackup gui are no longer available from
the background menu. Because the
Keck Observatory Archive (KOA) now ingests all raw data from every
instrument at the observatory, observers are encouraged to use KOA to retrieve
a copy of the data following your observing
sessions. Instructions and links to possible methods of data
backup including KOA, scp, and rsync are found at:
BackingUp.
DRP updated
(2015-Aug-3)
An updated version of the MOSFIRE DRP is released for observer
use. For details please see
https://keck-datareductionpipelines.github.io/MosfireDRP/ .
This release provides a number of new capabilities and
improvements as well as bug fixes. It is also highly enhances the
automation of the reduction process. Among the new pipeline
capabilities are:
- Automatic generation of the driver file
- Reduction of longslit observations
- reduction of long2pos and long2pos_specphot
- detailed log file and log output
- K-sigma clipping during wavelength fitting
- A bypass mode to autofit for a wavelength solution
- Package installation similar to the ureka package
- support for Numpy 1.9
MOSFIRE Drift corrected
(2015-Apr-29)
- MOSFIRE observers detected that objects drifted
along the slit over time. With the help of the MOSDEF team
and Dr. Chuck Steidel, we determined that the observed drift was
due to flexure in the guiding system and differential
atmospheric refraction.
Software updates were applied to both MOSFIRE and MAGIQ (the
guider). MAGIQ now corrects for the flexure in the system by
adjusting the guide star slightly as the instrument rotates and
the telescope changes elevation. MOSFIRE software now provides
MAGIQ with values for the filter central wavelength and ensures
that MAGIQ DAR corrections are enabled.
On sky testing during two nights appeared to show that
objects appear to remain in the same location with their
assigned slits. FITS headers will also indicate whether DAR is
enabled and the central wavelength used for correction.
Telescope Offset Problem fixed
(2014-Dec-15)
- We have addressed a long standing observatory wide problem
where commands to offset or nod the telescope were lost in
the command queries to the telescope. In the past, the
telescope would occasionally not nod to the desired
location, resulting in two consecutive spectra with the objects at
exactly the same location spatially in the slit. So a nod
sequence may look like this: A,B,A,B,A,B,B,C,B,C,B,C.
To solve this problem we implemented a gateway keyword
server. This is relatively transparent to observers. Most
importantly, since installing the gateway server in July 2014,
we no longer observe a skipped telescope offset. All nod
command work as expected. This solution was deployed to all
instruments at the observatory.
This issue was initially reported in April 2014 and was a
problem for all observations until the 2014B semester.
MOSFIRE DRP now supported by Keck
(2014-Dec-15)
- The MOSFIRE DRP is now being served and maintained by the
Keck Observatory. With this change there are some minor updates
in the latest release which should be transparent to DRP
users. A new issue tracking system is also implemented to help
us better track issues, questions, and concerns. Please see
https://mosfire-datareductionpipeline.github.io/MosfireDRP/ .
Drift in spectroscopic exposures
(2014-May-15)
- Observers are noticing a 1 pixel/hour drift along the slit
while acquiring spectroscopic observations with MOSFIRE. The
graph below shows the steady motion of a slit star in a mask
observed by the MOSDEF team. Over a two hour observing
sequence using a 4 point slit dither pattern, the star
drifted by 2 pixels.
Working with Dr. Richard Ellis, Matt Schenker was the first
to report the drift which was later confirmed by members
of the MOSDEF team, Dr. Mariska Kriek in particular.
The drift does not appear to be present in all
spectroscopic observations as some observations show a
maximum drift of 1 pix per hour while other observations
indicate that the star was stationary in the slit. We are
currently investigating if the drift is correlated with
telescope parameters such as elevations, azimuth, rotator
position angle (physical drive angle), etc. We have
exonerated the guiding software during engineering
completed in mid-May.
The MOSDEF team and Dr. Ellis's observing group monitor
the drift by designing their mask with a suitably bright
enough star that may be detected in a single nod pair of
observations. The star is bright enough that the teams can
monitor the position and intensity of the star. Keck
observatory suggests that observers who are concerned
about slit drift include a relatively bright star
comparable to those used for alignment (15-19 mag) in the
science mask configuration. The magnitude of the star will
depend on the length of individual exposures.
We are in the process of also porting over software used
by the teams to help monitor the position of the star in
real time so that observers may determine if a re-alignment is
necessary.
Telescope Offset Monitoring software released
(2014-Apr-14)
- On occasion during a spectroscopic nod sequence, the
telescope does not nod to the desired location. This will
often result in two consecutive spectra with the objects at
exactly the same location spatially in the slit. We have
recently released software that monitors the requested
positions and confirms telescope motion. If a move is
suspected of not executing, TKlogger, will pop up a message
on your display screen suggesting a course of action.
The typical course of action is to abort the dither
sequence and then restart the sequence. However, the
software is in its infancy. It is in the observer's best
interested to confirm that the telescope did not move by
examining the last image and the current image to asses if
the targets offset as expected or are at the same location
spatially in the slit. Please talk to your SA, if you have
questions.
DVD writer for data backups will be phased out
(2014-Mar-7)
- The Keck Observatory is phasing out the auto DVD writer
sometimes used by observers for data backups. Although there
is no time line for the removal of the DVD writer, we will no
longer troubleshoot or maintain the DVD writer. Because the
Keck Observatory Archive (KOA) now ingests all raw data from every
instrument at the observatory, observers are encouraged to use KOA to retrieve
a copy of the data following your observing
sessions. Instructions and links to possible methods of data
backup including KOA, scp, and rsync are found at:
BackingUp.
MOSFIRE DRP updated
(2014-Mar-7)
- Dr. Nick Konidaris has updated the DRP code to improve
maintenance, eradicate some persistent bugs, and address
some problems that observers like you have documented either
via e-mail or at the the google docs repository
forum.
The updated pipeline may be found at:
http://www.astro.caltech.edu/~npk/mosdrp_releases/
Major changes include:
- MOSDRP now has a user manual. Please offer
suggestions for improvement.
- Can combine arclamps and skylines for wavelength
calibration at the long end of K.
- Uses arclamps for long slit spectra.
- Fixes issues with integration times and now spits
out integration time maps for each reduced output.
- Provides standard deviation maps instead of ivar maps
- Divides the observations into
maskname/date/band. It does not attempt to combine
observations across multiple nights and now does not
have the attendant bugs with this step.
- Uses *Ureka* Python instead of STScI Python. Python
is now easy to install on Mac and linux (perhaps on
windows too but I haven't tested it).
- Uses .tar.gz files that are tagged by date, this
cleans up the versioning of the pipeline and doesn't
require users install Mercurial anymore.
Flat Lamps Updated
(2014-Feb-13)
- New flatfielding lamps installed atop the HIRES and AO
enclosures in Keck I replace the previous MOSFIRE dome lamp
attached to the Nasmyth Deck. The old flat lamp restricted
telescope access by Keck personnel. With the new lamps in use,
the telescope no longer needs to be set to an azimuth of 7°;
instead, the telescope and dome only need to be 90° away
from each other. The MOSFIRE calibration GUI is now updated to
use the new lamp, and includes revised default exposure
times. Spectrophotometric observers using extra-wide slits will
find that the standard exposure times on the calibration GUI
(which are intended for typical 0.7 arcsec slits) will
saturate the detector. Instead, observers with wide slits
should acquire their calibrations manually (i.e., not using the
Calibration GUI) and should use the new Domelamp Control GUI to
reduce the power level of the lamp. With low-power (i.e., high
attenuation) settings, spectrophotometric observers can obtain
flats with the lamps on and better illuminate the spectra at all
wavelengths. Click the image below to see the full-sized flat
lamp gui.
To launch the new GUI via the MOSFIRE workspace menu,
right-click on the background and select
K1 Telescope Status Menu > NEW Dome Lamp Control GUI
New Observing Mode Gui
(2014-Feb-5)
- The Observing Mode GUI was updated with a new layout. The
new version removes unnecessary options and adds new methods
for configuring to a dark filter. The new "quick dark"
option efficiently moves the filter wheel to the nearest
filter wheel combination to create a dark. Click on the
image below to see the full-size version of the new
ObsModeGui.
MOSFIRE DRP
(2013-Dec-16)
- Keck has revised instructions for installing the DRP in
conjunction with the newly releases Ureka Python package
supported by Gemini and STSCI. Please see: Keck DRP install instructions for the procedure. The official DRP code repository resides at http://code.google.com/p/mosfire
PIG Unleashed
(2013-Dec-04)
- We have added the new Program Interface
GUI (PIG) widget to the MOSFIRE user interface. On split nights,
observers use the PIG to indicate which observing program to
associate with the current image. This permits the Keck
Observatory Archive (KOA) to tag the images with the
appropriate observing program and observer list, and ensures
that the images are assigned the correct proprietary period.
MAGMA coordinate output feature
(2013 Dec 1)
- A post mortem of a failed slitmask alignment revealed that
MAGMA has a feature in the mask center coordinates. Recently
a mask was designed with the Declination seconds of arc
listed as "60.00" in the mask design output files (mask.xml,
starlist, etc.). Although the telescope DCS handled 60.000
arcsec the Slitmask Alignment Tool DSS query failed,
resulting in a difficult coarse alignment. Please check your
coordinate files to ensure proper formatting. Keck staff
will update software to check the coordinates (SAT and
MAGMA) in the near future.
New MOSFIRE FAQ Page May Answer
Your Questions
(2013 Sep 29)
- We have compiled answers to a number of common MOSFIRE
questions on the new MOSFIRE FAQ page.
'scent' Command Permits Reverse Moves
(2013 Jun 21)
- The scent command is used to offset the
telescope such that a target at image location (x,y) moves to
the slit center. However, sometimes observers may want to do
the opposite operation; i.e., move a target from the slit center
to a given pixel position. This is now possible by supplying
the -r (reverse) flag when invoking
scent. For example, an observer can place
a star at pixel (1000,500) as follows:
- Have the OA center the star on the guider at the REF
pointing origin.
- Have the OA send the target to the SLIT
pointing origin.
- Execute the command
scent -r 1000 500
This technique is useful for putting a star on a wide (10
arcsec) slit for spectroscopy.
dp Command Displays Images to ds9
(2013 May 31)
- The new dp command allows you to easily display
MOSFIRE images on the ds9 image display, and will overplot the
predicted slit positions if the image was a direct image. The
syntax is quite flexible:
- dp alone will display the most recent image
on disk
- dp n will display image n
on disk
- dp -n will display the
n-th previous image
- dp name.fits will display the
named image. If a relative pathname is supplied, the
image may be in either the current directory or the
current data directory.
This command is available on all MOSFIRE accounts when logged
into the MOSFIRE host computer, nuu.
Guider Communication Compromized
(2013 April 24 )
- The fiber optic cable on MOSFIRE's interconnect panel was
damaged in the morning of 22 April. Fiber cables are used to
communicate to the instrument and guider with the guider
requiring two sets of fibers. The guider fibers were damaged
preventing communication with the guider. All other
instrument control behaved normally. After much
troubleshooting the guider is in a usable state after
troubleshooting the issue for two days. We now have full
control over the guider detector but we have to manually
focus the guider during the night. This will cause some
minor delays at the beginning of the night. We plan to
resolve the issue after MOSFIRE come off sky in early May.
Calibration Tool Updates
(2013 April 5 )
- The GUI and scripts used to automate calibration acquisitions for
user designed slitmasks was updated and released for
observer use. The new GUI (seen below) presents a couple of
new options to the observer by giving them options for
changing arc-lamp exposure times and acquiring a flat lamp off
image of the dome. Default values for exposure times and
quantities were updated. The acquisition script is much
more intelligent and now ensures that masks properly execute, skips masks if no
exposures are requested, and is in general, more robust to
errors in mechanism moves. Like DEIMOS and LRIS observers,
MOSFIRE observers should be able to setup their calibration
sequence in the afternoon, initiate the script, and safely
grab dinner.
CSU Fatal Error Avoidance Strategies II: The New Voodoo
(2013 Feb 10 )
- The configurable slitmask unit (CSU) has been experiencing
“Fatal Errors” when reconfiguring to new
masks. This can occur when you are changing fields or when you
are reconfiguring from an alignment to a science mask. There
are new strategies to avoid a fatal error. First, all
afternoon calibrations are should be acquired with the instrument at a PA
of -90°. Second, before changing masks,
ensure that the drive angle on FACSUM is at least 30°
from zero. If the drive angle is within 10° of zero or
-180, please ask the
OA to rotate MOSFIRE 30° from the present location and
then initiate your mask move. Note that the drive angle is
likely different than the current PA of MOSFIRE. Both angles
are located in the bottom right on FACSUM.
TRUITIME vs ITIME
(2013 Jan 30 )
- Observers who require knowing the exposure time to a high
precision should note that there are two FITS header
keywords that specify the exposure time.
- TRUITIME - this is the true exposure time and has
units of seconds. This exposure time is within one read
time of the ITIME value below.
- ITIME - This is the exposure time as it was
specified in the gui or on the command line using tint.
CSU Fatal Error Avoidance Strategies
(2012 Nov 15 )
- The configurable slitmask unit (CSU) has been experiencing
"Fatal Errors" when reconfiguring to new masks. This can
occur when you are changing fields or when you are
reconfiguring from an alignment to a science mask. There are
two strategies for observing that may mitigate the CSU
errors that are recommended by the MOSFIRE team.These two
strategies apply when there are significant rotator or
telescope moves.
- First method developed by Ian McLean (UCLA)
- Put the dark filter in beam.
- Slew to the new field.
- Power cycle the CSU (your SA can instruct you).
- Send the mask setup.
- Execute the mask.
- Second method employed by Chuck Steidel (Caltech)
- Put the dark filter in beam.
- Slew to the new field.
- While slewing send the mask setup.
- Re-send the mask setup, when telescope is in
position and guiding.
- Execute the mask.
In addition to the night time observing strategies
above, we also recommend power cycling the CSU just before
starting your calibrations in the afternoon.
New recovery software has
reduced the lost time to 20-25 min.most of the
time. Less frequently, fatal errors may result in up
to 1.5 hours of lost time.
MAGMA - slitmask design software released
(2012 Aug 7 )
- MAGMA is the software used to design slitmask for
MOSFIRE. Version 1.0 was released today and may be down
loaded for use following the links at
MAGMA . A version of the software is available for
MAC, Linux, Solaris, and Windows. We are still shaking out
a few bugs, so if you observe an issue, please contact Marc
Kassis.
MOSFIRE scheduled for 2012B
(2012 June )
- MOSFIRE is scheduled for regular use starting in
2012B, and all commissioning efforts are expected to be
completed before the first night of observing in
September. For observers, who are scheduled to use MOSFIRE
in 2012B, you will need to prepare slitmask design files in
advance of your run. The new slitmask design software is
called MAGMA and will be released in early August. Please
see the MOSFIRE observatory
newsletter article for first light details.
MOSFIRE First Light
(2012 April 4)
- MOSFIRE first light images were
acquired on the night of April 4, despite heavy circus
clouds over Mauna kea. Images and spectra were acquired on
the night of the 5 as part of the continuing commissioning
of the instrument. Please check the web pages for
information on the performance of the instrument. The
observatory press release may be found at
observatory news
MOSFIRE moved to RT1
(2012 Feb 22)
- After complete unpacking from the shipping crate, MOSFIRE
was lifted to the RT1 position on the Keck I telescope
Nasmyth Deck next to LRIS.
MOSFIRE arrives at Keck!
(2012 Feb 16)
- MOSFIRE arrived safely at the Keck Observatory early
afternoon on 16 February 2012 after traveling by boat from
the mainland. MOSFIRE was uncrated and placed inside the
dome for unpacking.