The DEIMOS internal focus can change significantly on a daily
basis; hence, it's advisable to refocus the spectrograph each
afternoon. Since the offsets between various filters appear
to be stable over time, the standard procedure is to focus the
instrument in direct imaging mode using the R filter
and dome illumination. Once the focus in this fiducial filter
is measured and set, the instrument's automated filter focus
discipline software will adjust the internal focus as needed
to account for the changes in the filter. This document
describes the procedure for measuring the fiducial focus.
Acquisition
Focus is measured using dome illumination through the grid of
holes masks in direct imaging mode with the R-band
filter. Follow these steps to acquire a focus sequence in this
mode:
- Verify that the telescope is released on the Instrument
Activity Schedule. Telescope status should be set to
- "Ready"
- with green background color
- Confirm telescope pointing. On FACSUM verify that:
- telescope elevation is 45°
- telescope azimuth value is 90°
different from the dome azimuth
- Ensure that the rotator is unlocked:
- click the Rotation button on
the Dashboard to view the Rotator subpanel.
- If there is a button reading UNLOCK in red
lettering on a black background, click it to unlock the
rotator.
- From the window manager pulldown menu, select the item
DEIMOS Utilities > Run/analyze focusloop (R+dome)
This will configure the spectrograph appropriately for R-band
dome images, acquire a sequence of 7 focus images every 1000
focus units from -5000 to -11000, analyze the results, and
display a focus map and best focus value.
When finished, the script produces a plot and displays it
on your screen in a ghostview window.
- Inspect the plot to verify that the overall morphology of
the contours is similar to that seen in this sample plot.
- If the plot looks good, adopt the last value shown
on the page or in the xterm output (labeled Array
focus) as the nominal focus value. Proceed to
Setting the Focus below.
- If it looks bad (see example of a buckled mask resulting in a
poor focus like
this), try the following:
- Rotate DEIMOS to a different rotator position (DEIMOS
PA=-30 is most likely to succeed).
- Click the Retract Arm button on the
slitmask panel to remove the GOH mask.
- Re-select the GOH mask from the main Dashboard panel.
- Take a single image using the internal lamps and
check for incredibly out of focus images in the slitmask
regions which look unreasonable on the plot.
- Once the image appears to be in reasonable focus
across the entire mask, re-run the focusloop procedure.
- Proceed to Setting the Focus below
Setting the focus
- On the Dashboard GUI, enter the nominal focus value in
the Focus entry blank in the filter box.
- Double-click in the filter box to raise the filter detail
panel.
- Change the filter from R to anything else
- Verify that the values of:
- FOCUSOFF is non-zero
- FFD Status is Ready.
Return to the Afternoon
Checklist.