DEIMOS CCD Controller Powerdown Procedure | |
BackgroundThis document descibes the procedure for turning off power to the DEIMOS science CCD array or the Flexure Compensation System (FCS) detector system.Procedure
TranscriptsScience CCD SystemHere is a transcript of a successful completion of the DEIMOS powerdown command on the science CCD crate:**** WARNING **** CYCLING POWER ON THE CCD CONTROLLER WILL INJECT CHARGE INTO THE CCDS. THIS INJECTED CHARGE WILL TAKE 10 TO 20 HOURS TO DECAY AFTER POWER ON. YOU SHOULD AVOID POWERING DOWN THE CONTROLLER IF YOU ARE PLANNING TO OBSERVE WITH THIS CCD WITHIN THE NEXT 24 HOURS. ARE YOU SURE YOU WANT TO CONTINUE WITH THIS POWER DOWN SEQUENCE? (yes/no) yes Will proceed with power down sequence in 10 seconds Waiting 60 seconds for power down sequence to complete The CCD controller for service deiccd has been powered down It is now safe to turn off the main power switch on the front panel of the DEIMOS science CCD controller power supply chassis that is located in bay 8 of the DEIMOS electronics ring FCS CCD SystemHere is a transcript of a successful completion of the DEIMOS powerdown command on the FCS CCD crate:**** WARNING **** CYCLING POWER ON THE CCD CONTROLLER WILL INJECT CHARGE INTO THE CCDS. THIS INJECTED CHARGE WILL TAKE 10 TO 20 HOURS TO DECAY AFTER POWER ON. YOU SHOULD AVOID POWERING DOWN THE CONTROLLER IF YOU ARE PLANNING TO OBSERVE WITH THIS CCD WITHIN THE NEXT 24 HOURS. ARE YOU SURE YOU WANT TO CONTINUE WITH THIS POWER DOWN SEQUENCE? (yes/no) yes Will proceed with power down sequence in 10 seconds Powering down DEIMOS CCD controller service deifcs Waiting 60 seconds for power down sequence to complete The CCD controller for service deifcs has been powered down It is now safe to turn off the main power switch on the front panel of the DEIMOS FCS CCD controller chassis that is located in bay 10 of the DEIMOS electronics ringIf the powerdown sequence is not successful, the following message will be shown in the terminal: Waiting 60 seconds for power down sequence to complete The +35 volt analog power for this controller did not turn off The +/- 15 volt analog power for this controller did not turn off The analog switches in the dewar did not open as commanded *** The power down sequence for DEIMOS CCD controller service deifcs *** was not successful. Please call UCO/Lick at UC Santa CruzTo power down the FCS CCD in an orderly manner when the command powerdown deifcs fails, modify the following keywords: modify -s deifcs utbn2fil=off modify -s deifcs utb30ven=disabled modify -s deifcs utb15ven=disabled modify -s deifcs utbn2htr=offYou can check that the system was powered down correctly by running again the commabnd powerdown deifcs. The output should be as follows: The CCD controller for service deifcs has already been powered down | |
Last modified: 07/10/2020 05:20 Send questions or comments to:DEIMOS Support |