An accurate measurement of the time of the flux-weighted midpoint of each exposure is essential to correct for the barycentric velocity of the Earth. During an exposure, a small fraction of the light within the spectrometer is diverted to a separate, low-resolution spectrometer that records the flux time series for all wavelengths across the main spectrometer bandpass. This system accurately tracks the photon arrival times in the parent spectrometer, allowing for precise determination of flux-weighted exposure mid-points as a function of wavelength.

The optical design of the exposure meter is a prism-based spectrometer with a resolving power of approximately 100. The exposure meter accepts light from two fibers; one collecting light from the otherwise unused outboard slices of the science fiber at the spectrometer reformatter entrance, and a sky fiber from the fiber injection unit on the telescope. The optical design is shown in the following figure.

Last modified: 06/21/2023 06:59
Send questions or comments to:KPF Support

The information on this page is the property of the W. M. Keck Observatory. The contents of this page or any part thereof shall not be copied or otherwise reproduced or transferred to other documents or used or disclosed to others for any purpose other than observing support at the W. M. Keck Observatory and the subsequent analysis and publication of scientific data obtained from observations conducted at the W. M. Keck Observatory. All rights reserved. © W. M. Keck Observatory.