Instrument Issues
DRP Issues
Flux Misassignment -- Not really misassignment
What Can I do?
What is Keck Doing?

Dated Entries

2022 Jan 31: Imager Flats Available

2021 Apr 14: ETCs & Strehl Estimator.

2021 Mar 12: DRP v5.0.0 Released

2020 Sep 22: DRP v5.0beta Released

2018 May 07: DRP v4.2.0 Released

Mar 30, 2018 -- OSIRIS IMAG UPGRADE --

Apr 18, 2016 -- OSIRIS SPEC UPGRADE -- -- NOTES

Jan 26, 2016 -- DVD writer no longer available for data backups

The Keck Observatory is phasing out the auto DVD writer sometimes used by observers for data backups. Options to launch the DVD autobackup gui are no longer available from the background menu. Because the Keck Observatory Archive (KOA) now ingests all raw data from every instrument at the observatory, observers are encouraged to use KOA to retrieve a copy of the data following your observing sessions. Instructions and links to possible methods of data backup including KOA, scp, and rsync are found at: BackingUp.

November 5, 2015 -- Updated Recmats

We have updated the current rectification matrices (recmats) to use a median combination of calibration scans. In most cases, this dramatically reduces the incidents of extremely large absolute value pixel values.

Rectification Matrices

June 25, 2014 -- Modified Startup Scripts

Users will note additional software is started automatically when selecting via the pull-down menu:

OSIRIS Control Menu --> Start OSIRIS Software

April 3, 2014 -- New IDL legend command breaks scaled sky subtraction

Symptom
Scaled sky subtraction stops working
Problem
IDL added a legend command in v8.3 that supersedes the astro library command of the same name
Solution
  1. Download new IDL astro library or simply rename your current IDL astro legend.pro to al_legend.pro and fix the PRO call within the file
  2. Change line 851 of scaledskysub_000.pro from:
    legend, /top, /right, $
    to
    al_legend, /top, /right, $

March 7, 2014 -- DVD writer for data backups will be phased out

The Keck Observatory is phasing out the auto DVD writer sometimes used by observers for data backups. Although there is no time line for the removal of the DVD writer, we will no longer troubleshoot or maintain the DVD writer. Because the Keck Observatory Archive (KOA) now ingests all raw data from every instrument at the observatory, observers are encouraged to use KOA to retrieve a copy of the data following your observing sessions. Instructions and links to possible methods of data backup including KOA, scp, and rsync are found at: BackingUp.

September 16, 2013 -- 2012 Recmats Available

Recmats from 2012 needed to be reprocessed to take advantage of fixes in DRP v3.2. They are available for download.

August 1, 2013 -- OSIRIS DRP v3.2 Released

Download it today!

New 2013 rectification matrices are available. New 2012 recmats are forthcoming.

We recommend all users with data after September 2011 re-reduce their data with v3.2 of the DRP and new rectification matrices. We found that one of the video cards is misplacing a column of data in the raw files.

January 1, 2013 -- OSIRIS Grating Upgrade

We have successfully upgraded the grating in OSIRIS. Initial on-sky measurements confirm an increase in throughput by a factor of 2!

January 17, 2012 -- OSIRIS Moved to Keck I

Craned off Keck II, rolled down to Keck I.

January 12-13, 2012 -- OSIRIS Service Mission

We repaired the lenslet mask stage, adjusted the focus positions of the SPEC scales to attemp parfocality, and round a clearance issue with the 50 mas scale.

January 1, 2012 -- More targets in the new year!

In the spirit of giving this holiday season, US Space Command recently increased the number of LGSAO targets that observers are permitted to submit for LGSAO observations. The new maximum per night is 128 total targets. For split nights, each half night observing team is allocated 64 targets.

November 18, 2011 -- Unplanned warmup

Due to a reconfig error, OSIRIS warmed unexpectedly on November 10. OSIRIS is now cold again, but there is a possibility that the rectification matrices may need recalibration. With the current status of the lenslet mask (broken), we cannot take calibrations now.

September 21, 2011 -- Broadband Filters Only

Yesterday the OSIRIS lenslet mask stopped working correctly. This is the stage that couples with the filter wheel when switching between broadband and narrowband filters. The stage is currently in the broadband position and any attempts to move the lenslet mask to the narrowband position fail. This appears to be a hardware issue inside the dewar. OSIRIS with broadband filters is currently available and working correctly.

August 22, 2011 -- Persistence Characterization

The SPEC detector shows persistence on exposures as low as 500 total data numbers. More information is available on the persistence and saturation page.

July 2, 2010 -- New Calibrations Needed

We have recently fixed an alignment issue with the K2AO calibration unit that caused the OSIRIS calibration scans since the October 2009 service to suffer from spatial and spectral fringing. This means you will need new rectification matrices and re-reduce your data. We are working on re-taking all the OSIRIS calibrations. This is a lengthy process that will likely take until the end of September. Please contact if you have any questions

April 6, 2010

To support the release of DRP v2.3, we have made new rectification matrices by reprocessing all calibrations scans taken after the October 2009 service. The new calibration pipeline has corrected many issues attributed to rectifcation matrices; however, there are still some "bad spaxels". We are spot checking matrices and attempting to correct them. Please contact if you see "bad spaxels" in your reductions.

New versions of rectificaion matrices from the warm period of OSIRIS are listed according to SPEC detector temperature (keyword SCURTMP).

OSIRIS Rectification Matrices

March 8, 2010

DRP v2.3 released. Here is a message from the OSIRIS team:

Dear OSIRIS users,

We have just released the latest version (v2.3) of the OSIRIS pipeline. We recommend that you use this new pipeline for OSIRIS data taken after January 2009. A number of the pipeline changes (listed below) were made in order to solve issues that arose during OSIRIS' warmer temperature period. We recommend unpacking and installing this new version in a separate directory to existing versions on your computer. You may download v2.3 OSIRIS pipeline and Mac Os X install script at the following link.
OSIRIS Pipeline Download Page

We have also modified the Calibration Reduction Pipeline (CRP) to generate adjusted and cleaner calibration files used during 'Extract Spectra' for data taken after January 2009. I have posted a document that outlines some issues with previous calibration files on the OSIRIS wiki (linked under "OSIRIS Calibration or Rectification Matrices"). We recommend using calibration files generated from v2.3 for all data taken from January 2009 to present-day.
OSIRIS Rectification Matrices

The OSIRIS pipeline is 100% supported by donated time and goodwill. To maintain the high quality of OSIRIS reductions requires all of our contributions. Modifications and testing of this new version (v2.3) of the pipeline was aided by a number of individuals within the community. Special thanks goes to Tuan Do, Quinn Konopacky, Jessica Lu, Nicholas McConnell, Marshall Perrin, and Jonelle Walsh. There is a new OSIRIS v2.3 manual that documents these changes. Please use the OSIRIS wiki page (link above) to post issues with this pipeline and your own solutions! :)
Happy reductions!
Sincerely, Shelley, James, and OSIRIS team

December 9, 2009

OSIRIS Rectification Matrices Updated. These will calibrate data after the October 2009 Server that returned OSIRIS to normal temperatures.

December 7, 2009

Imager wheel #1 seems to fault everytime we move it. It seems stuck in its home position (Hn1). Darks are still possible by moving imager wheel #2 to a different filter such that the filters are crossed (e.g. Hn1 + Kn2). There is no actual dark slide for the OSIRIS imager.

December 4, 2009

For all OSIRIS observers during summer 2009, new rectification matrices will be coming. We have identified an issue in the calibration frames and are working on a fix. We appreciate your patience.

November 17, 2009

The OSIRIS service mission was a success. OSIRIS is at normal operating temperatures.

Quick SPEC characteristics:
Temp (K) Dark Current
900s Dark (DN/s)
Dark Current
900s Dark (e-/s)
Read Noise
900s Dark (e-)
Time for
DC=RN (s)
Measurements
67.7±0.1 1.29±0.07E-02 5.62±0.29E-02 12.0±0.8 213.5 45

October 28, 2009

The OSIRIS cooldown has started. Initial indications are positive and this cooldown looks more normal than other 2009 cooldowns. Temperature and pressure plots are available on the temps page. The comparison cooldown is the successful March 2008 cooldown.

October 14, 2009

We serviced OSIRIS over the weekend. We strongly suspect that we had a poor thermal contact between the cold head and copper block. This would produce a thermal bottleneck inside the instrument. We think that we have improved the contact, but time will tell. We replaced a thermal shield that was damaged and removed during a prior service. OSIRIS is pumping down. Cooldown will start in about 1 week.

During the service, we also measured the grating efficiency. We are analyzing data and will deliver a report to the observatory and SSC. Because the grating was removed during the service, a complete set of calibrations will be necessary.

September 23, 2009

We will service OSIRIS during October to further investigate the cooling issues. We plan to have OSIRIS back online by mid-November.

August 26, 2009

The OSIRIS cold head was successfully exchanged for a new unit. It appears that our OSIRIS cooling issues are still with us. We are letting the instrument cool to see how cold it will get.

July 30, 2009

OSIRIS dewar temperatures were rising at 2-3 K/day and accelerating. OSIRIS has been taken offline and allowed to warm for a servie to replace the cold head.

July 29, 2009

We are no longer able to control the SPEC detector temperature. It is above 84 K and rising at about 0.1 K/hr. Dark current and read noise are both elevated; however, pairwise subtracted darks show no unusual structure. OSIRIS will likely be unusable in the next few days.

June 15, 2009

We have found 75.5 K to be a stable temperature for SPEC. Sensitivity is down 10-15% due to the increased dark current, but otherwise performance has been ok. In its new state, the spatial and spectral response has changed in the following ways:

Spatial rectification
New Rectification Matrices (v5.0) are available for most observed modes. If you have observed another mode since March 2009 for which a rectification matrix is unavailable, please let us know right away.
Spectral rectification
Due to the dewar temperature increase of about 10 K, the wavelength calibration has changed by about 2 Angstroms. We are working on a new wavelength calibration.

May 18, 2009

Warm temperatures on the summit have helped push the SPEC temerature higher. Recent operating temperatures have been 72 K, then 73 K, then 74 K, 75 K, and 75.5 K. I will continue taking darks to monitor performance.

April 20, 2009

OSIRIS is usable, but operating warmer than before. Temperatures within the dewar are 8 to 10 K above normal. We have stabilized the SPEC detector to 72 K, 4 K above our nominal 68 K operating temperature. Dark frames look good cosmetically, subtract well, and seem stable over several days. The IMAG detector remains at its nominal 70 K.

Warmer temperature darks are available on our public ftp site.

Temp (K) Mean Counts
900s Dark (DN/s)
Median Counts
900s Dark (DN/s)
Measurements
68.0±0.1 1.64±0.74E-02 1.47E-02 208
72.0±0.1 2.08±0.23E-02 2.06E-02 8
73.0±0.1 2.34±0.21E-02 2.30E-02 16
74.0±0.1 2.60±0.13E-02 2.55E-02 29
75.0±0.1 3.15±0.20E-02 3.09E-02 17
75.5±0.1 3.32±0.19E-02 3.29E-02 8
78.0±0.1 4.47±0.19E-02 4.47E-02 8
79.0±0.1 5.45±0.08E-02 5.45E-02 5
80.0±0.1 6.50±0.14E-02 6.53E-02 10
80.0±0.1 6.79±0.19E-02 6.77E-02 26
83.5±0.1 1.28±0.005E-01 1.28E-01 5
84.5±0.1 1.57±0.01E-01 1.57E-01 7

The wide range for dark current at 68 K is likely due to not selecting data. That is, some darks were taken immediately after the detector was reset, which causes a known noise artifact.

Read Noise using the "2-bias method": RN = stdev(im1-im2) / sqrt(2) / gain
where: im1 and im2 are darks and gain = 0.23 DN/e

900 second Dark Frames and Subtractions. All images have Histogram Equilization stretch. Please note ranges.

T=73 K Dark T=72 K Dark
T=68 K Dark

Subtraction of 2 T=73 K Darks (same day) Subtraction of 2 T=72 K Darks (different days)
Subtraction T=72 Dark from T=68 K Dark

February 20, 2009

OSIRIS has suffered from thermal issues that have caused higher dark current in the SPEC detector. We have also seen images shift when changing plate scales. We have recently completed a service of OSIRIS and reached the following conclusions:

OSIRIS was serviced in February 2009. We think that we have repaired a thermal short, but we were unable to determine a cause of the image motion with changing plate scale. It's possible that the thermal cycle will change the performance of the SPEC image motion. We expect to return OSIRIS to full capability on March 31, 2009.

For more details, please read our Preliminary Report.

May 9, 2008

The following is a email sent to all known OSIRIS users regarding the March 2008 service mission.

Dear OSIRIS users,

There have been significant changes to OSIRIS since March 2008 and we want you to know about the changes. If you have received this email in error, please accept my apologies.

  1. In early March we serviced OSIRIS to add new K-band filters with built-in pupils for the 100 mas scale. These new filters are called Kcb, Kc3, Kc4, and Kc5 and are identical to the Kbb, Kn3, Kn4, and Kn5 filters respectively. Both sets of K-band filters are in the instrument. To make room for these new filters, we removed the Open, Zn2, Zn3, and Zn5 filters. The new KcX filters are meant to be used only with the 100 mas plate scale. For all other plate scales, please continue to use the original K-band filters. We have engineering time in June to measure the difference in background between the old and new filters. All operational software has been updated to reflect the changes. Observers may wish to download new versions of the Manual, OOPGUI, and DRP from the OSIRIS Toolbox web page.
  2. During this mission, we fixed a mechanical issue with the OSIRIS handling cart that prevented the instrument from defining in the proper location. Since the mission, we can now measure the focus offset among the 4 SPEC plate scales and the Imager.
  3. We have developed software to automatically correct for the focus offsets among the 4 SPEC plate scales and the imager. This software is transparent to the observer, but there are important things to note:
    1. If you are taking simultaneous SPEC and IMAG frames, sky frames will be focused for the imager, object frames will be focused for the spectrograph.
    2. If you are taking SPEC frames only, object and sky frames will be focused for the SPEC plate scale.
    3. If you are taking IMAG frames only, object and sky frames will be focused for the Imager.
  4. We have converted OSIRIS over to use the kvnc suite of software developed for mainland observing. OSIRIS will no longer use the osiris server as its VNC server; rather, it will use svncserver2. The change should be transparent for Waimea observers and mainland observers will no longer have to hand-edit a script to start the proper VNC viewer.
Please contact with any questions and thank you!

jim

April 28, 2008

OSIRIS will now automatically account for the focus offset among the scales when scripts are run via the GUIs.

April 28, 2008

The OSIRIS Rectification Matrices Webpage has been updated with the new "Kc" filters under the "v4.0" heading. Except for the new "Kc" filters, all other matrices are identical to "v3.0".

April 28, 2008

Links for the OSIRIS Manual and for software on the Tools Page have been updated.

March 7, 2008

OSIRIS was serviced yesterday by Keck and UCLA personnel. We had two goals for the service: investigate and correct the global and relative focus jumps; install new K-band filters for the 100 mas scale.

We found that the OSIRIS handling cart was stopping too far from the AO bench. This seemed to prevent the instrument from defining correctly. We adjusted the switch on the handling cart to allow it to move closer to the AO bench.

Inside the dewar, we found the locking mechanism on the lens barrels for the 35- and 100-mas scales to be loose. Unexpectedly, we found the 50-mas lens barrel to be secure. We tightened the locking mechanism for all scales and adjusted the 50-mas to more closely align to the 20-mas scale, which is our reference point. OSIRIS is still warm so testing has not started.

We removed the Opn, Zn2, Zn3, and Zn5 filters and replaced them with filters equivalent to the Kbb, Kn3, Kn4, and Kn5 filters. These new filters also have 9-m pupils installed that will significantly reduce the background observed in the 100-mas scale. These new filters are named Kcb, Kc3, Kc4, and Kc5, where the 'c' is for coarse. These filters are intended only for the 100 mas scale. Software has been updated and tested.

March 5, 2008

Recent fiber testing with OSIRIS has confirmed that the 4 OSIRIS plate scales are no longer parfocal. This accounts for part of the image quality problems observers have seen in the 50 mas scale. An OSIRIS service mission starts tomorrow which will attempt to correct the problem.
Scale (mas) Focal shift (mm)
20 0
35 -0.25 ± 0.25
50 -1.75 ± 0.25
100 0

August 21, 2007

Recent fiber testing with OSIRIS has confirmed that the 4 OSIRIS plate scales are no longer parfocal. This accounts for part of the image quality problems observers have seen in the 50 mas scale. Most likely, the 2006 earthquake or subsequent service mission caused the focal shift. We are investigating ways to correct or compensate for the focal shift.
Scale (mas) Focal shift (mm)
20 0
35 0.50 ± 0.25
50 1.00 ± 0.25
100 0

August 2, 2007

The OSIRIS rectification matrices are now online and available for download.

August 1, 2007

Jim Lyke has produced a DRP Cookbook to help users get started in reducing data.

July 26, 2007

Recently, we've noticed some image quality issues related to the 50 mas scale of OSIRIS. We are investigating the cause. Possiblities include:
  1. We have an incorrect offset to the center of the 50 mas scale such that we are vignetting the beam slightly.
  2. There is a focus offset for the 50 mas scale.
  3. Something on the AO bench changes with OSIRIS spec scale

In a possibly related note, some observers have recently described problems in offsetting from the center of one scale to another. This could indicate an incorrect offset value for the 50 mas scale.

We need the AO bench to investigate this issue. OSIRIS will next be in AO between August 5 - 22, 2007.

June 6, 2007 note to observers from James Larkin

Dear OSIRIS users,

We have just released the latest version of the pipeline, quicklook2, and OSIRIS manual (version 2.0). There have been significant improvements and changes to the pipeline, we highly recommend that all users refer to the manual for your reductions. In addition we have included a new interactive data reduction pipeline gui (ODRFGUI) which allows users to easily operate and generate pipeline files. This new GUI should be run with the latest version of the pipeline (version 2.0). All of this can be downloaded from the

Keck OSIRIS Toolbox or the UCLA OSIRIS webpage.

IMPORTANT: It was discovered that all calibration scans taken in March 2006 for K and H broadband and narrowband 100 mas scale modes were saturated. If you had observations with these modes from March 2006 - May 2007 please contact the for new calibration files for your reductions. For more information please see the new OSIRIS Updates page.

List of Major Changes to the Pipeline

  1. Added new module "Correct Dispersion" - corrects for atmospheric dispersion and chromatic dispersion from the AO bench and should be performed on all OSIRIS cubes after "Assemble Data Cube" in the final reduction processes (GUI puts this in the correct order).
  2. Added a new module "Extract Star" - extracts 1D spectrum of a stellar object from an OSIRIS cube, typically for telluric extraction.
  3. Added a new module "Remove Hydrogen Lines" - takes a 1-D spectrum and attempts to remove absorption lines due to hydrogen. The primary purpose is to remove hydrogen absorption lines from telluric standard stars before dividing them into science exposures.
  4. Added a new module "Divide by Blackbody" - divides a 1-d, 2-d or 3-d spectrum by a blackbody of a specified effective temperature (Kelvin). This is primarily useful to fix telluric spectra, and accepts different wavelength ranges.
  5. Added a new module "Divide by Star Spectrum" - Divides cube by 1-d stellar spectrum. This is primarily useful for telluric correction.
  6. Fixed wavelength solution to resolve small ~0.1 A shifts between each lenslet wavelength solution.
  7. The "Mosaic Frames" module now updates the RA and DEC header in the output file when multiple frames are mosaiced together.

Happy Reductions!

Best,

James

Feb 27, 2007

An updated version of the OSIRIS Manual v1.1 (PDF - 1.8 MB) is available.

Feb 14, 2007

OSIRIS was damaged in the October 15, 2006 Earthquake. One of the G-10 A-frames that support the primary optical plate broke, causing a thermal touch-off within the dewar. James Larkin and our team at Keck opened OSIRIS in late December. We diagnosed and fixed the broken A-frame. Subsequent calibration and on-sky data have shown that the alignment of the optical elements within OSIRIS is the same as before the earthquake. There appears to be no change to the wavelength calibration nor to the spatial recitifcation of data.

Feb 22-23, 2005

OSIRIS achieves first light! The 22/23 February run is the first of 5 runs planned for the instrument commissioning. In the following months, all the instrument modes and operations with AO, both with Natural Guide Star and Laser Guide Star, will be tested.

OSIRIS Installation on Keck II