Using engineering data acquired in Aug and Sept, piston
corrections using the offset guider IQM for LRIS was
calibrated. As of 25 Oct. 2009, IQM is released for use with MIRA
observations. The intention is to demonstrate whether IQM
suggests telescope focus adjustments that are comparble to
those determined using the MIRA software.
IQM setup procedure
- Slew to an appropriate MIRA star
- Run MIRA and SEND
CORRECTIONS
- Select the REF0 as the PO and run a MAGIQ camera focus
routine. From the oA GUI top menu bar select Camera->Focusing
- Select from DCSGUI/MAGIQ menu Start IQM UI to
lanch the IQM screens
- From the IQM UI
- Confirm that on the IQM UI
- Select lrisOffset from pull-down camera list and press "Select" button
- Confirm that it is selected
- Measurement Only checkbox is
selected.
- IQM center set to "Click"
- Click One Shot on IQM UI
- Click on the star in the image to acquire IQM data
- Check that the coordinates beside the IQM center show
the ones close to where you clicked.
IQM acquisition procedure: T01040
- Slew to an appropriate MIRA star
- Select REFO as the PO and center the star
- Select the ND1 filter for the offset guider. Photometric filter should be set to clear.
- Confirm that on the IQM UI
- lrisOffset is selected and click 'Select'
- Measurement Only checkbox is
selected.
- IQM center set to "Click"
- Click One Shot on IQM UI
- Click on the star in the image to acquire IQM data
- Check that the coordinates beside the IQM center show
the ones close to where you clicked.
- Wait until the full frame is back.
- Run MIRA and SEND CORRECTIONS
- Repeat steps 2-7
- Continue with science obs.
IQM sample screen
IQM UI screen grab with IQM image loaded. The three red arrows
at left indicate where needed options are to be set.