LRIS Grating Changeout Procedure

Only Trained Personnel should change Gratings

Changing A Grating

1.      Login to honolii as lriseng.

2.      Select 'Engineering | Changeout | All Reconfiguration Software' in the desktop menu

3.      Click on the grating position to be changed from the GUI or from a lrisserver xterm prompt  type 'gr_service x' to drive the grating to the service port. Where x is the number of the grating to be changed, (i.e. gr_service 5, to rotate the grating in position 5 to the service port).

4.      LRIS must be removed to the Cass Platform to change a grating.  Remove LRIS following the module servicing procedure.

5.      Locate the nut tray, a grating cover and the yellow handled 5/16 nut driver in the gray cabinet on nas deck.

6.      Rotate LRIS to the grating change out position with the dewars at 12 o'clock (180�).

7.      Set up step ladder on the right side of LRIS to gain access to the service port.

8.      Open the service port cover.  Slide the nut trap under the grating cell to prevent the grating retaining nuts from falling into the turret and to prevent the turret from being inadvertently rotated.

9.      Install the grating cover securely.

10.  Remove the nuts retaining the grating into the cell with the nut driver.

11.  Remove the grating from the turret by pulling on the cover handles, being careful not to shock the grating cell and stow in the gray cabinet.

12.  Install the required grating in the turret.

13.  Install the nuts by putting the nut in the socket of the nut driver first, then screw to tighten.  Sometimes the gratings cells do not define until they are screwed in snugly.  Remove the grating cover and nut trap.

14.  Drive the turret to the next position to be serviced and install the grating.

15.  Close and latch the service port cover.

16.  Stow the nut driver, grating cover, and nut trap in the gray cabinet on the nas deck.

17.  Five gratings should remain in the cabinet.  Five gratings stay in LRIS.

Reconfiguring Software Table to Reflect the Change

1.        Use the LRIS Reconfiguration GUI.

2.      Repeat until all grating that were installed are reconfigured.

3.      If the graphical tool fails you can reconfigure using low-level commands.

4.      Verify the changes by using the Optic Configuration GUIs from the pull down and hit update or by typing "consort" on lrisserver xterm to see new reconfiguration.

If There Are No Other LRIS Configuration Changes...

1.      Install LRIS in the telescope as per procedures.

2.      Select 'Engineering | Checkout LRIS' in the desktop menu.

3.      Press 'y' to start the checkout script.

4.      Note if the script has any problems executing and notify support astronomer or inst tech.

5.      Select Shut Down from workspace 'Engineering | LRIS Control Menu | Shutdown LRIS s/w'.  This should be done to ensure background tasks are closed and HQ can start up LRIS.

6.      Select logout from the desktop menu.

7.      Update the summit activities instrument schedule on the web.

Doc Revision History

  • 1999 Mar 18 - Eli Morris
  • 2004Mar21 - scd - Updated for in situ changeout, GUI control and proper shutdown
Last modified: 12/28/2011 20:17
Send questions or comments to:LRIS Support

The information on this page is the property of the W. M. Keck Observatory. The contents of this page or any part thereof shall not be copied or otherwise reproduced or transferred to other documents or used or disclosed to others for any purpose other than observing support at the W. M. Keck Observatory and the subsequent analysis and publication of scientific data obtained from observations conducted at the W. M. Keck Observatory. All rights reserved. © W. M. Keck Observatory.