The proper focus of each camera varies with filter and observing mode (e.g., direct imaging and slitmask spectroscopy vs. longslit spectroscopy) and, for the red side, with the dichroic in use. Even though the zero point for the focus changes daily, the relative offsets between various focus modes on LRIS are believed to be stable. Hence, if the observer focuses the instrument in the nominal imaging mode then the focus for all other instrument modes can be derived from that measurement.
This document describes the officially-endorsed procedure for daily focusing of LRIS. The procedure involves using a pinhole mask in the LRIS focal plane to obtain a series of images with varying focus values, then analyzing these images to derive the optimal focus for imaging mode. Once the instrument is focused, the observer generates a table showing the corresponding focus values in other modes.
Note: if you are NOT going to do any imaging, you can focus the blue side only, and forget about the red side. This saves quite a lot of time.
LRIS Utilities > Run Imaging Focusloops > DOME lamps, blue & red (DEFAULT)Change this to blue only if you are focusing the blue camera only.
Press <Enter> to exit...Press the Enter key to terminate this window.
LRIS Utilities > Xfocus > Xfocus red
LRIS Utilities > Xfocus > Xfocus blue
|
|
LRIS Utilities > Print Focus TableThe focus table will be printed on the printer in Remote Ops I.
LRIS Utilities > Save Focus Table
LRIS Utilities > View Focus Table
focus_tableor print it by executing either
focus_table | lpr -Plw4
focus_table | lpr -Plw6
This table is only valid if you are starting in a configuration known
to be in good focus. Typicaly you would run xfocus in some
configuration, set to the recommended focus and then run "focus_table".
The current setup is read, tabulated offsets calculated from the
red and blue filters, red and blue focus values, the dichroic, and
whether there is a long slit in or not. Then new values are calculated
for all other combinations.
R E D S I D E F O C U S
Filter/Grism Imaging/multislit Longslit
=============================================================
mirror -0.660 -0.620
150/7500 -0.644 -0.604
300/5000 -0.623 -0.583
400/8500 -0.662 -0.622
600/5000 99.340 99.380
600/7500 -0.690 -0.650
600/10000 -0.649 -0.609
831/8200 -1.020 -0.980
900/5500 -0.690 -0.650
1200/7500 -0.663 -0.623
clear -0.700
B 99.340
V -0.825
R -0.660
Rs -0.659
I -0.700
GG495 -0.667
OG570 -0.589
RG850 -0.700
B L U E S I D E F O C U S
Filter/Grism Imaging/multislit Longslit
=============================================================
clear -3443 -3523
300/5000 -3425 -3505
600/4000 -3415 -3495
400/3400 -3403 -3483
1200/3400 -3357 -3437
U -3423
B -3463
G -3443
V -3408
NB4170 -3588
NB4040 -3370
NB4430 -3409
NB3875 -3501
NB3985 -3417
NB3640 -3433
NOTE: There are no dichroic offsets for the red or blue side.
Focus values of order 100 indicate no offset has been measured for that optic.