The dispersive elements include grisms available on the Blue side and gratings
that are available on the Red side. The grisms on the blue side are always
installed and available to the observers. The gratings on the red side are
available to observers, but need to be specified on the instrument configuration
form to ensure that the desired gratings are installed during an observing
session.
The table below lists the grisms available for use with LRIS on the
Blue Side only
The values presented were empirically measured. Chuck Steidel (Caltech)
supplied the wavelength ranges and efficiency data. The FWHM of
lines using a 1 arcsec slit width were obtained using the internal
lamps, and ranges in FWHM are from blue to red in wavelength. The
efficiencies on the blue side are the efficiency of the instrument
with that grism.
Grism
|
Dispersion
|
Efficiency |
Wavelenth Range (Å)
|
Flat Field |
FWHM resolution for 1" slit (A) |
MOS
| Longslit
|
300/5000 |
1.43 |
data
 |
2240-8090
| 1600-7450 |
|
8.4-9.2 |
400/3400 |
1.09 |
data
 |
1790-6250 |
1300-5770 |
|
6.5-7.1 |
600/4000 |
0.63 |
data
 |
3330-5910 |
3040-5630 |
|
3.8-4.1 |
1200/3400 |
0.24 |
data
 |
3040-4030 |
2940-3920 |
|
1.56 |
Note: The efficiency data was acquired by Chuck Steide (4 June
2002). The efficiency plots reflect the efficiency of LRISb with each
grism. For the 300, 400, & 600 grisms, the mirror was used at the
position of the dichoric. The UV response improves with the use of
dichroics instead of the mirror. The 1200 grism efficiency is in
combination with the 460 dichroic. All other optics were the same for
all grism observations. Below is a comparison plot of the instrument throughputs for the
four grisms as measured by Chuck Steide (4 June 2002)
The following gratings are available for use with LRIS red side. The
grating information is up to date for the red side upgrade. Dispersion
and wavelength coverage are calculated and are not measured
values. The efficiency plots are not up to date and were determined
using the old detector.
Grating
|
Dispersion
|
Efficiency1 |
Δλ (Å) |
Flat Field |
FWHM resolution for 1" slit (Å) |
150/7500 |
3.0 |
|
12288 |
n/a
|
300/5000 |
1.59 |
|
6525 |
n/a
|
9.18 |
400/8500 |
1.16 |
 |
4762 |
n/a
|
6.9 |
600/5000 |
0.80 |
 |
3275 |
 |
4.7 |
600/7500 |
0.80 |
 |
3275 |
 |
4.7 |
600/10000 Gold coated |
0.80 |
 |
3275 |
 |
4.7 |
831/8200 Gold coated |
0.58 |
 |
2375 |
n/a
|
900/5500 |
0.53 |
 |
2175 |
 |
1200/7500 |
0.40 |
 |
1638 |
 |
1200/9000 Gold coated |
0.40 |
|
1638 |
TBD
|
Mirror |
n/a |
 |
n/a |
n/a |
1. The grating efficiency was measured by
acquiring slitless spectra of spectrophotometric standard stars. The
grating, standard star, date of observation, and filter(s) used are indicated on
the plots. The spectra were reduced as follows:
- Bias subtraction of stellar spectrum and calibration images.
- Flat-fielding of stellar and arc-lamp spectrum images; the
flat-field images were taken with internal illumination, generally at
the same sky-PA as the stellar spectrum.
- Extraction of stellar spectra with background subtraction; no
variance weighting is applied but the spectra have S/N typically
greater than 50. The atmospheric A and B-O2 bands, and the H-alpha
absorption line when appropriate, are removed.
- The extracted spectra are extinction corrected and integrated
over bandpasses from the STScI "medium"-resolution tables (2-10 Å
spacing). Then, the spectra are fluxed against the tabulated
brightness values.
- The resulting efficiencies are smoothed over 50 Å (typical)
centers, and plotted as postscript files.
- The plots for the 600/10000 and 1200/9000 were
derived in 2022.
Additional plots for the red side
In addition to the efficiency plots in the table above, several
comparative plots between gratings of similar resolution are presented
to illustrate crossover points and relative efficiencies.
Notes:
- Several spectra
that have overlapping free spectral ranges at the same resolution
display an efficiency reduction at the blue end of each progressively
red spectrum; see 1200g/mm for example.This effect must be due to a
reduced transmission of the CCD over some range of columns on the left
side of the CCD (see also Grating flat-field plots).
- Several of the
curves (150/7500, 600/7500, and 831/8200) display "enhanced"
efficiencies near 6500 Å; in the case of 600/7500 it is seen also in
the flat-field plot and is apparently due to a grating ghost. In
addition 1200g/mm has a ghost at 4500 Å of similar nature.
- More data exists beyond 9200 A but these tables end there. The spectrum
400g/mm-8500 Å blaze with GG495 was not flat-fielded.