Index
Hardward description and basics
The blue side temperature controller is a Lakeshore 336.
This is the manual.
This controller is dedicated to the control of the CCD temperatures, using a closed loop controlled by input A with output on output2.
Configuration
Inputs |
A |
Blue CCD temperature |
|
B |
Blue Camera Barrel |
|
C |
Blue Camera Shutter |
|
D1 |
Invar plate |
|
D2 |
Blue Cold Strap bot |
|
D3 |
Blue Cold strap top |
|
D4 |
Blue CCD case |
|
D5 |
Lower bench |
|
A |
B |
C |
D |
D2 |
D3 |
D4 |
D5 |
Input Name |
CCD temperature |
Camera Barrel |
Camera Shutter |
Invar plate |
Cold Strap Bot |
Cold Strap Top |
CCD Case |
Lower Bench |
Sensor Type |
PTC TRD (Platinum) |
Diode |
Diode |
Diode |
Diode |
Diode |
Diode |
Diode |
Autorange |
On |
2.5 V (Silicon) |
2.5 V (Silicon) |
2.5 V (Silicon) |
2.5 V (silicon) |
2.5 V (silicon) |
2.5 V (Silicon) |
2.5 V (Silicon) |
Current reversal |
On |
|
|
|
|
|
|
|
Curve |
32 |
35 |
34 |
33 |
33 |
33 |
33 |
35 |
Filter |
Off |
Off |
Off |
Off |
Off |
Off |
Off |
Off |
Temperature Limit |
375 K |
0 K |
0 K |
350 K |
350 K |
350 K |
350 K |
0 K |
Preferred units |
Kelvin |
Kelvin |
Kelvin |
Kelvin |
Kelvin |
Kelvin |
Kelvin |
Kelvin |
Diode current |
|
1 mA |
1 mA |
10 microA |
10 micro A |
10 micro A |
10 micro A |
1 mA |
|
32 |
|
Ohms |
Kelvin |
Name |
KCWI CCD rtd |
1 |
16.4865 |
50 |
Serial number |
kcwirtd |
2 |
27.48 |
77.35 |
Format |
Ohms/Kelvin |
3 |
37.089 |
100 |
Setpoint Limit |
375.000 K |
4 |
57.888 |
150 |
Temp Coefficient |
Positive |
5 |
78.111 |
200 |
|
|
6 |
97.945 |
250 |
|
|
7 |
117.454 |
300 |
|
|
8 |
155.74 |
400 |
|
|
9 |
0 |
0 |
|
33 |
|
Volts |
Kelvin |
|
23 User kcwilabdio |
1 |
0.298 |
333 |
Name |
KCWI LAB DIODE |
2 |
0.401 |
294.8 |
Serial number |
kcwilabdio |
3 |
0.404 |
293 |
Format |
Volts/Kelvin |
4 |
0.997 |
77.35 |
Setpoint Limit |
375.000K |
5 |
0 |
0 |
Temp Coefficient |
Negative |
|
|
|
|
|
|
Volts |
Kelvin |
|
35 |
1 |
0.488 |
325 |
Name |
Cam Barrel diod |
2 |
0.587 |
294.5 |
Serial number |
|
3 |
0.654 |
273.9 |
Format |
Volts/Kelvin |
4 |
0.732 |
250 |
Setpoint Limit |
0 K |
5 |
0 |
0 |
Temp Coefficient |
Negative |
|
|
|
|
34 |
|
Volts |
Kelvin |
Name |
Cam shutter diod |
1 |
0.4 |
381.7 |
Serial number |
|
2 |
0.5 |
272.7 |
Format |
Volts/Kelvin |
3 |
0.6 |
163.8 |
Setpoint Limit |
375 |
|
|
|
Temp Coefficient |
Negative |
|
|
|
OUTPUT SETUP |
|
|
1 - not active |
2 |
|
|
Output mode |
Closed loop PID |
Closed loop PID |
|
|
Control input |
D4: Input D4 |
A |
|
|
Heater Resistance |
25 Ohms |
25 Ohms |
|
|
Max Current (25ohm) |
0.707 A (12.5 W) |
0.707 A (12.5 W) |
|
|
Power Up Enable |
Off |
On |
|
|
Heater Out Display |
Power |
Power |
|
|
SetPoint Ramping |
Off |
Off |
Curves and settings for upload
To upload these settings and the curves, you can use a browser connected to the private network, and go to http://lakeshore, then use the Utilities
Location of sensors