The blue side temperature controller is a Lakeshore 336.
This is the manual.
This controller is dedicated to the control of the CCD temperatures, using a closed loop controlled by input A with output on output2.
Inputs | A | CCD temperature |
B | Camera Barrel | |
C | Camera Shutter | |
D1 | Inval plate | |
D2 | Cold Strap bot | |
D3 | Cold strap top | |
D4 | CCD case | |
D5 | Lower bench |
A | B | C | D | D2 | D3 | D4 | D5 | |
Input Name | CCD temperature | Camera Barrel | Camera Shutter | Invar plate | Cold Strap Bot | Cold Strap Top | CCD Case | Lower Bench |
Sensor Type | PTC TRD (Platinum) | Diode | Diode | Diode | Diode | Diode | Diode | Diode |
Autorange | On | 2.5 V (Silicon) | 2.5 V (Silicon) | 2.5 V (Silicon) | 2.5 V (silicon) | 2.5 V (silicon) | 2.5 V (Silicon) | 2.5 V (Silicon) |
Current reversal | On | |||||||
Curve | 32 | 35 | 34 | 33 | 33 | 33 | 33 | 35 |
Filter | Off | Off | Off | Off | Off | Off | Off | Off |
Temperature Limit | 375 K | 0 K | 0 K | 350 K | 350 K | 350 K | 350 K | 0 K |
Preferred units | Kelvin | Kelvin | Kelvin | Kelvin | Kelvin | Kelvin | Kelvin | Kelvin |
Diode current | 1 mA | 1 mA | 10 microA | 10 micro A | 10 micro A | 10 micro A | 1 mA |
P | 25 | for output2 |
I | 5 | |
D | 5 |
32 | Ohms | Kelvin | ||
Name | KCWI CCD rtd | 1 | 16.4865 | 50 |
Serial number | kcwirtd | 2 | 27.48 | 77.35 |
Format | Ohms/Kelvin | 3 | 37.089 | 100 |
Setpoint Limit | 375.000 K | 4 | 57.888 | 150 |
Temp Coefficient | Positive | 5 | 78.111 | 200 |
6 | 97.945 | 250 | ||
7 | 117.454 | 300 | ||
8 | 155.74 | 400 | ||
9 | 0 | 0 |
33 | Volts | Kelvin | ||
23 User kcwilabdio | 1 | 0.298 | 333 | |
Name | KCWI LAB DIODE | 2 | 0.401 | 294.8 |
Serial number | kcwilabdio | 3 | 0.404 | 293 |
Format | Volts/Kelvin | 4 | 0.997 | 77.35 |
Setpoint Limit | 375.000K | 5 | 0 | 0 |
Temp Coefficient | Negative |
Volts | Kelvin | |||
35 | 1 | 0.488 | 325 | |
Name | Cam Barrel diod | 2 | 0.587 | 294.5 |
Serial number | 3 | 0.654 | 273.9 | |
Format | Volts/Kelvin | 4 | 0.732 | 250 |
Setpoint Limit | 0 K | 5 | 0 | 0 |
Temp Coefficient | Negative |
34 | Volts | Kelvin | ||
Name | Cam shutter diod | 1 | 0.4 | 381.7 |
Serial number | 2 | 0.5 | 272.7 | |
Format | Volts/Kelvin | 3 | 0.6 | 163.8 |
Setpoint Limit | 375 | |||
Temp Coefficient | Negative |
OUTPUT SETUP | 1 - not active | 2 | ||
Output mode | Closed loop PID | Closed loop PID | ||
Control input | D4: Input D4 | A | ||
Heater Resistance | 25 Ohms | 25 Ohms | ||
Max Current (25ohm) | 0.707 A (12.5 W) | 0.707 A (12.5 W) | ||
Power Up Enable | Off | On | ||
Heater Out Display | Power | Power | ||
SetPoint Ramping | Off | Off |
File | |
32 | XLS |
33 | XLS |
34 | XLS |
35 | XLS |
Overall config | ICG |
To upload these settings and the curves, you can use a browser connected to the private network, and go to http://lakeshore, then use the Utilities