11.1. CuAr Calibration

Here is an example of step-by-step instructions for obtaining a wavelength calibration spectrum in the echelle mode.

  1. Start the ESI software.

  2. On the GUI, select Echelle mode, the slit width of choice, and Clear_S in the lower filter wheel. Select Full Frame readout from the QuickFrame button. (Alternatively load the setup file for the echelle slit of your choice.)

  3. Select the CuAr lamp from either the menu or the arc lamp panel. (Remember to hit "Move" if you use the menu.)

  4. Make sure the hatch is closed, and set the exposure time in the ``CCD'' panel to 300 seconds. (With the hatch closed and a lamp on, a red beam will run from the hatch all the way to the CCD panel in the GUI.)

  5. Click ``Move'' to set the stages, and ``UpdateCCD'' to set the CCD parameters.

  6. Enter the object name annd click the ``Expose'' button.

11.2. Obtaining a Target Spectrum in the Echelle Mode

  1. Start the ESI software.

  2. Select Observing Mode of Echelle, Clear_S for the lower wheel, and your slit width from the Slit panel. Select Full Frame readout from the QuickFrame button. (Alternatively, load the setup file appropriate for the echelle slit of your choice.)

  3. Make sure the hatch cover is open, and enter your desired exposure time. (A blue beam should now be visible in the GUI.)

  4. Click ``Move'' and ``UpdateCCD''.

  5. Enter the object name and click the Expose button.

Last modified: 12/30/2013 07:54
Send questions or comments to:ESI Support

The information on this page is the property of the W. M. Keck Observatory. The contents of this page or any part thereof shall not be copied or otherwise reproduced or transferred to other documents or used or disclosed to others for any purpose other than observing support at the W. M. Keck Observatory and the subsequent analysis and publication of scientific data obtained from observations conducted at the W. M. Keck Observatory. All rights reserved. © W. M. Keck Observatory.