If you need to manually power cycle the crates, the crates for both the science and FCS detectors are located in the K2 computer room (see figure 1 below)
|
|
start_crate f
rebootto bring the system up cleanly. This will produce the following message that can be ignored:
Timeout waiting for ISERV, retrying Got network configuration Broadcast message sent on ln0 Broadcast not sent on lo0
shutdownto verify that all systems are properly turned off.
< restartto invoke the restart script and verify that it completes without error.
^]
quit
[61] dmoseng@polo: start_crate fcs First, reboot the controller: -> reboot ... Next, determine what tasks are running: -> i NAME ENTRY TID PRI STATUS PC SP ERRNO DELAY ---------- ------------ -------- --- ---------- -------- -------- ------- ----- tExcTask _excTask ffed08 0 PEND 94950 ffeb20 d0003 0 tLogTask _logTask ffc3b0 0 PEND 94950 ffc1c8 0 0 tShell _shell fe8d58 1 READY 7f31c fe86a0 30065 0 tRlogind _rlogind ff4b00 2 PEND 7bdfc ff45a8 0 0 tTelnetd _telnetd ff2bf8 2 PEND 7bdfc ff2900 0 0 tRlogOutTas_rlogOutTask 6ff168 2 READY 7bdfc 6fed88 0 0 tRlogInTask_rlogInTask fdbac0 2 READY 7bd00 fdb678 0 0 tNetTask _netTask ff7fb8 50 READY 7cf64 ff7d00 0 0 tFtpdTask _ftpdTask ff0118 55 PEND 7bdfc fefe10 0 0 ccdClock _ccdClock e3b528 90 READY 7e974 e3b3c0 0 0 tPortmapd _portmapd ff1688 100 PEND 7bdfc ff1330 16 0 MLOG_STDOUT_start_mlog_ e6c638 100 PEND 94950 e6c0a8 0 0 cserv _cserv e53db0 100 PEND 7bdfc e537a8 0 0 broad_mon _broadcast_m e159e8 100 DELAY 7e974 e15880 3d0001 18 responder _responder e130f0 100 PEND 94950 e11d88 0 0 rccd _rccd e2d1d0 150 PEND+T 7bdfc e2ced8 3d0004 18 value = 0 = 0x0 Turn off system in an orderly manner: -> shutdown Have entered shutdown routine Setting exit flags for monitor processes Waiting for monitor processes to shutdown Sending STP command to TIMING board Sending LOR command to TIMING board RD bias voltages have all been set to minimum voltage Analog switches in dewar electronics commanded open Analog switches in dewar electronics indicate open Analog power has been commanded off Now switch off main power at controller value = 0 = 0x0 Next, determine what tasks are running: -> i NAME ENTRY TID PRI STATUS PC SP ERRNO DELAY ---------- ------------ -------- --- ---------- -------- -------- ------- ----- tExcTask _excTask ffed08 0 PEND 94950 ffeb20 d0003 0 tLogTask _logTask ffc3b0 0 PEND 94950 ffc1c8 0 0 tShell _shell fe8d58 1 READY 7f31c fe86a0 30065 0 tRlogind _rlogind ff4b00 2 PEND 7bdfc ff45a8 0 0 tTelnetd _telnetd ff2bf8 2 PEND 7bdfc ff2900 0 0 tRlogOutTas_rlogOutTask 6ff168 2 READY 7bdfc 6fed88 0 0 tRlogInTask_rlogInTask fdbac0 2 READY 7bd00 fdb678 0 0 tNetTask _netTask ff7fb8 50 READY 7cf64 ff7d00 0 0 tFtpdTask _ftpdTask ff0118 55 PEND 7bdfc fefe10 0 0 ccdClock _ccdClock e3b528 90 READY 7e974 e3b3c0 0 0 tPortmapd _portmapd ff1688 100 PEND 7bdfc ff1330 16 0 MLOG_STDOUT_start_mlog_ e6c638 100 PEND 94950 e6c0a8 0 0 cserv _cserv e53db0 100 PEND 7bdfc e537a8 0 0 responder _responder e130f0 100 PEND 94950 e11d88 0 0 rccd _rccd e2d1d0 150 PEND+T 7bdfc e2ced8 3d0004 39 value = 0 = 0x0 Launch the restart script: -> < restart startup("/vx/force5ce/deimos/vme.deifcs/hosts/dsp",0) Sending hardware reset to VMEINF board Resetting VMEINF-2 by writing 0 to 6efff004 interrupt: Unknown level 15 Interrupt! Sending STP command to TIMING board Sending LOR command to TIMING board RD bias voltages have all been set to minimum voltage UTB application 0 test location= 10 UTB digital output= ff80 Analog switches in dewar electronics commanded open Analog switches in dewar electronics indicate open Sending hardware reset to TIMING board msg size = 2 words. reply size = 2 words msg = 0X102 0X54544852 reply = 0X10002 0X444f4e Sending STP command to TIMING board Will verify vmeinf eprom against /vx/force5ce/deimos/vme.deifcs/hosts/dsp/vmeinf/vme.s VMEINF board verification status= 0, should be 0 Will verify timing eprom against /vx/force5ce/deimos/vme.deifcs/hosts/dsp/timing/timing.s TIMING board verification status= 0, should be 0 Will verify utility eprom against /vx/force5ce/deimos/vme.deifcs/hosts/dsp/utility/util.s Expecting 1 verification error: at 800081 verify error: at address 800081 expected: 000004 got: 052d79 EOF UTILITY board verification status= 1, should be 1 EEPROMs verified correctly. Proceeding with power-up +5v readback raw ADC reading = 3413 volts= 4.991480 prior to enables +15v readback raw ADC reading = 3486 voltage = 15.971211 prior to enable -15v readback raw ADC reading = 616 voltage = -16.102915 prior to enable +30v readback raw ADC reading = 2048 volts = 0.000000 prior to enable UTB digital output= ff80 +/-15v power enabled +5v readback raw ADC reading = 3416 volts= 5.002450 after 15v enable +15v readback raw ADC reading = 3486 voltage = 15.971211 after 15v enable -15v readback raw ADC reading = 617 voltage = -16.091669 after 15v enable +30v readback raw ADC reading = 2048 volts = 0.000000 after 15v enable UTB digital output= fe80 +30v power enabled +5v readback raw ADC reading = 3416 volts= 5.002450 prior after enables +15v readback raw ADC reading = 3487 voltage = 15.982318 after enables -15v readback raw ADC reading = 617 voltage = -16.091669 after enables +30v readback raw ADC reading = 3284 volts = 34.679161 after enables Analog switches in dewar electronics commanded open Analog switches in dewar electronics indicate open Bias and clock voltages have been set, but not yet sent to CCD Bias voltages for CCD 1 CCD 1 bias 0 ADC 3435 ( d6b) MUX 2.026 Volts 20.258 Vdd-1 CCD 1 bias 1 ADC 3435 ( d6b) MUX 2.026 Volts 20.258 Vdd-2 CCD 1 bias 2 ADC 2868 ( b34) MUX 1.196 Volts 11.959 RD-A CCD 1 bias 3 ADC 2868 ( b34) MUX 1.196 Volts 11.959 RD-B CCD 1 bias 4 ADC 1709 ( 6ad) MUX -0.498 Volts -0.996 OG-A CCD 1 bias 5 ADC 2047 ( 7ff) MUX -0.003 Volts -0.006 Spare-2 CCD 1 bias 6 ADC 2047 ( 7ff) MUX -0.003 Volts -0.033 Guard ring CCD 1 bias 7 ADC 2051 ( 803) MUX 0.003 Volts 0.003 Substrate CCD 1 bias 8 ADC 3390 ( d3e) MUX 1.961 Volts 3.921 Spare-1 CCD 1 bias 9 ADC 1709 ( 6ad) MUX -0.498 Volts -0.996 OG-B CCD 1 bias 10 ADC 2047 ( 7ff) MUX -0.003 Volts -0.006 Spare-3 CCD 1 bias 11 ADC 2047 ( 7ff) MUX -0.003 Volts -0.029 Bi-polar Spare CCD 1 bias 12 ADC 2047 ( 7ff) MUX -0.003 Volts 0.000 Ground CCD 1 bias 13 ADC 2047 ( 7ff) MUX -0.003 Volts 0.000 Ground CCD 1 bias 14 ADC 2047 ( 7ff) MUX -0.003 Volts 0.000 Ground CCD 1 bias 15 ADC 2047 ( 7ff) MUX -0.003 Volts 0.000 Ground Bias voltages for CCD 2 CCD 2 bias 0 ADC 3436 ( d6c) MUX 2.027 Volts 20.273 Vdd-1 CCD 2 bias 1 ADC 3436 ( d6c) MUX 2.027 Volts 20.273 Vdd-2 CCD 2 bias 2 ADC 2868 ( b34) MUX 1.196 Volts 11.959 RD-A CCD 2 bias 3 ADC 2868 ( b34) MUX 1.196 Volts 11.959 RD-B CCD 2 bias 4 ADC 1709 ( 6ad) MUX -0.498 Volts -0.996 OG-A CCD 2 bias 5 ADC 2047 ( 7ff) MUX -0.003 Volts -0.006 Spare-2 CCD 2 bias 6 ADC 2048 ( 800) MUX -0.002 Volts -0.018 Guard ring CCD 2 bias 7 ADC 2051 ( 803) MUX 0.003 Volts 0.003 Substrate CCD 2 bias 8 ADC 3388 ( d3c) MUX 1.958 Volts 3.915 Spare-1 CCD 2 bias 9 ADC 1709 ( 6ad) MUX -0.498 Volts -0.996 OG-B CCD 2 bias 10 ADC 2047 ( 7ff) MUX -0.003 Volts -0.006 Spare-3 CCD 2 bias 11 ADC 2047 ( 7ff) MUX -0.003 Volts -0.029 Bi-polar Spare CCD 2 bias 12 ADC 2047 ( 7ff) MUX -0.003 Volts 0.000 Ground CCD 2 bias 13 ADC 2047 ( 7ff) MUX -0.003 Volts 0.000 Ground CCD 2 bias 14 ADC 2047 ( 7ff) MUX -0.003 Volts 0.000 Ground CCD 2 bias 15 ADC 2047 ( 7ff) MUX -0.003 Volts 0.000 Ground Open bias voltages are all within tolerance Low-rail clock voltages for clock gen 1 CLKGEN 1 clock 0 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 1 clock 1 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 1 clock 2 ADC 1503 ( 5df) MUX -0.798 Volts -7.980 CLKGEN 1 clock 3 ADC 1708 ( 6ac) MUX -0.498 Volts -4.978 CLKGEN 1 clock 4 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 5 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 6 ADC 1691 ( 69b) MUX -0.523 Volts -5.227 CLKGEN 1 clock 7 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 8 ADC 1691 ( 69b) MUX -0.523 Volts -5.227 CLKGEN 1 clock 9 ADC 1707 ( 6ab) MUX -0.499 Volts -4.993 CLKGEN 1 clock 10 ADC 2047 ( 7ff) MUX -0.001 Volts -0.015 CLKGEN 1 clock 11 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 1 clock 12 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 1 clock 13 ADC 1503 ( 5df) MUX -0.798 Volts -7.980 CLKGEN 1 clock 14 ADC 1503 ( 5df) MUX -0.798 Volts -7.980 CLKGEN 1 clock 15 ADC 1708 ( 6ac) MUX -0.498 Volts -4.978 CLKGEN 1 clock 16 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 17 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 18 ADC 1691 ( 69b) MUX -0.523 Volts -5.227 CLKGEN 1 clock 19 ADC 1691 ( 69b) MUX -0.523 Volts -5.227 CLKGEN 1 clock 20 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 21 ADC 1707 ( 6ab) MUX -0.499 Volts -4.993 CLKGEN 1 clock 22 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 1 clock 23 ADC 2049 ( 801) MUX 0.001 Volts 0.015 High-rail clock voltages for clock gen 1 CLKGEN 1 clock 0 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 1 clock 1 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 1 clock 2 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 1 clock 3 ADC 2388 ( 954) MUX 0.498 Volts 4.978 CLKGEN 1 clock 4 ADC 2368 ( 940) MUX 0.469 Volts 4.685 CLKGEN 1 clock 5 ADC 2369 ( 941) MUX 0.470 Volts 4.700 CLKGEN 1 clock 6 ADC 2368 ( 940) MUX 0.469 Volts 4.685 CLKGEN 1 clock 7 ADC 2368 ( 940) MUX 0.469 Volts 4.685 CLKGEN 1 clock 8 ADC 2368 ( 940) MUX 0.469 Volts 4.685 CLKGEN 1 clock 9 ADC 2388 ( 954) MUX 0.498 Volts 4.978 CLKGEN 1 clock 10 ADC 2456 ( 998) MUX 0.597 Volts 5.974 CLKGEN 1 clock 11 ADC 2456 ( 998) MUX 0.597 Volts 5.974 CLKGEN 1 clock 12 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 1 clock 13 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 1 clock 14 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 1 clock 15 ADC 2389 ( 955) MUX 0.499 Volts 4.993 CLKGEN 1 clock 16 ADC 2368 ( 940) MUX 0.469 Volts 4.685 CLKGEN 1 clock 17 ADC 2369 ( 941) MUX 0.470 Volts 4.700 CLKGEN 1 clock 18 ADC 2368 ( 940) MUX 0.469 Volts 4.685 CLKGEN 1 clock 19 ADC 2369 ( 941) MUX 0.470 Volts 4.700 CLKGEN 1 clock 20 ADC 2368 ( 940) MUX 0.469 Volts 4.685 CLKGEN 1 clock 21 ADC 2388 ( 954) MUX 0.498 Volts 4.978 CLKGEN 1 clock 22 ADC 2456 ( 998) MUX 0.597 Volts 5.974 CLKGEN 1 clock 23 ADC 2456 ( 998) MUX 0.597 Volts 5.974 Low-rail clock voltages for clock gen 2 CLKGEN 2 clock 0 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 2 clock 1 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 2 clock 2 ADC 1503 ( 5df) MUX -0.798 Volts -7.980 CLKGEN 2 clock 3 ADC 1503 ( 5df) MUX -0.798 Volts -7.980 CLKGEN 2 clock 4 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 5 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 6 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 2 clock 7 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 2 clock 8 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 2 clock 9 ADC 1503 ( 5df) MUX -0.798 Volts -7.980 CLKGEN 2 clock 10 ADC 2046 ( 7fe) MUX -0.003 Volts -0.029 CLKGEN 2 clock 11 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 12 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 13 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 14 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 15 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 16 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 17 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 18 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 19 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 20 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 21 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 22 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 23 ADC 2048 ( 800) MUX 0.000 Volts 0.000 High-rail clock voltages for clock gen 2 CLKGEN 2 clock 0 ADC 2183 ( 887) MUX 0.198 Volts 1.977 CLKGEN 2 clock 1 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 2 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 3 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 4 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 5 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 6 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 7 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 8 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 9 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 10 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 11 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 12 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 13 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 14 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 15 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 16 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 17 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 18 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 19 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 20 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 21 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 22 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 23 ADC 2048 ( 800) MUX 0.000 Volts 0.000 Open clock voltages are all within tolerance Switch states set to select clock low rails RD bias voltages have all been set to minimum voltage Analog switches commanded closed. Analog switches in dewar electronics indicate closed +5v readback raw ADC reading = 3417 volts= 5.006107 after closure +15v readback raw ADC reading = 3486 voltage = 15.971211 after closure -15v readback raw ADC reading = 617 voltage = -16.091669 after closure +30v readback raw ADC reading = 3281 volts = 34.594990 after closure RD has been commanded to nominal voltage CCD is now powered up. Bias voltages for CCD 1 CCD 1 bias 0 ADC 3419 ( d5b) MUX 2.002 Volts 20.024 Vdd-1 CCD 1 bias 1 ADC 3419 ( d5b) MUX 2.002 Volts 20.024 Vdd-2 CCD 1 bias 2 ADC 2868 ( b34) MUX 1.196 Volts 11.959 RD-A CCD 1 bias 3 ADC 2868 ( b34) MUX 1.196 Volts 11.959 RD-B CCD 1 bias 4 ADC 1709 ( 6ad) MUX -0.498 Volts -0.996 OG-A CCD 1 bias 5 ADC 2048 ( 800) MUX -0.001 Volts -0.003 Spare-2 CCD 1 bias 6 ADC 2048 ( 800) MUX -0.002 Volts -0.018 Guard ring CCD 1 bias 7 ADC 2050 ( 802) MUX 0.001 Volts 0.001 Substrate CCD 1 bias 8 ADC 3390 ( d3e) MUX 1.961 Volts 3.921 Spare-1 CCD 1 bias 9 ADC 1709 ( 6ad) MUX -0.498 Volts -0.996 OG-B CCD 1 bias 10 ADC 2047 ( 7ff) MUX -0.003 Volts -0.006 Spare-3 CCD 1 bias 11 ADC 2047 ( 7ff) MUX -0.003 Volts -0.029 Bi-polar Spare CCD 1 bias 12 ADC 2047 ( 7ff) MUX -0.003 Volts 0.000 Ground CCD 1 bias 13 ADC 2047 ( 7ff) MUX -0.003 Volts 0.000 Ground CCD 1 bias 14 ADC 2047 ( 7ff) MUX -0.003 Volts 0.000 Ground CCD 1 bias 15 ADC 2047 ( 7ff) MUX -0.003 Volts 0.000 Ground Bias voltages for CCD 2 CCD 2 bias 0 ADC 3419 ( d5b) MUX 2.002 Volts 20.024 Vdd-1 CCD 2 bias 1 ADC 3419 ( d5b) MUX 2.002 Volts 20.024 Vdd-2 CCD 2 bias 2 ADC 2868 ( b34) MUX 1.196 Volts 11.959 RD-A CCD 2 bias 3 ADC 2868 ( b34) MUX 1.196 Volts 11.959 RD-B CCD 2 bias 4 ADC 1709 ( 6ad) MUX -0.498 Volts -0.996 OG-A CCD 2 bias 5 ADC 2048 ( 800) MUX -0.001 Volts -0.003 Spare-2 CCD 2 bias 6 ADC 2048 ( 800) MUX -0.002 Volts -0.018 Guard ring CCD 2 bias 7 ADC 2051 ( 803) MUX 0.003 Volts 0.003 Substrate CCD 2 bias 8 ADC 3388 ( d3c) MUX 1.958 Volts 3.915 Spare-1 CCD 2 bias 9 ADC 1709 ( 6ad) MUX -0.498 Volts -0.996 OG-B CCD 2 bias 10 ADC 2047 ( 7ff) MUX -0.003 Volts -0.006 Spare-3 CCD 2 bias 11 ADC 2047 ( 7ff) MUX -0.003 Volts -0.029 Bi-polar Spare CCD 2 bias 12 ADC 2047 ( 7ff) MUX -0.003 Volts 0.000 Ground CCD 2 bias 13 ADC 2048 ( 800) MUX -0.001 Volts 0.000 Ground CCD 2 bias 14 ADC 2048 ( 800) MUX -0.001 Volts 0.000 Ground CCD 2 bias 15 ADC 2048 ( 800) MUX -0.001 Volts 0.000 Ground Closed bias voltages are all within tolerance Low-rail clock voltages for clock gen 1 CLKGEN 1 clock 0 ADC 1505 ( 5e1) MUX -0.795 Volts -7.950 CLKGEN 1 clock 1 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 1 clock 2 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 1 clock 3 ADC 1708 ( 6ac) MUX -0.498 Volts -4.978 CLKGEN 1 clock 4 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 5 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 6 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 7 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 8 ADC 1691 ( 69b) MUX -0.523 Volts -5.227 CLKGEN 1 clock 9 ADC 1707 ( 6ab) MUX -0.499 Volts -4.993 CLKGEN 1 clock 10 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 1 clock 11 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 1 clock 12 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 1 clock 13 ADC 1503 ( 5df) MUX -0.798 Volts -7.980 CLKGEN 1 clock 14 ADC 1503 ( 5df) MUX -0.798 Volts -7.980 CLKGEN 1 clock 15 ADC 1707 ( 6ab) MUX -0.499 Volts -4.993 CLKGEN 1 clock 16 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 17 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 18 ADC 1691 ( 69b) MUX -0.523 Volts -5.227 CLKGEN 1 clock 19 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 20 ADC 1692 ( 69c) MUX -0.521 Volts -5.212 CLKGEN 1 clock 21 ADC 1708 ( 6ac) MUX -0.498 Volts -4.978 CLKGEN 1 clock 22 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 1 clock 23 ADC 2049 ( 801) MUX 0.001 Volts 0.015 High-rail clock voltages for clock gen 1 CLKGEN 1 clock 0 ADC 2185 ( 889) MUX 0.201 Volts 2.006 CLKGEN 1 clock 1 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 1 clock 2 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 1 clock 3 ADC 2388 ( 954) MUX 0.498 Volts 4.978 CLKGEN 1 clock 4 ADC 2369 ( 941) MUX 0.470 Volts 4.700 CLKGEN 1 clock 5 ADC 2368 ( 940) MUX 0.469 Volts 4.685 CLKGEN 1 clock 6 ADC 2368 ( 940) MUX 0.469 Volts 4.685 CLKGEN 1 clock 7 ADC 2369 ( 941) MUX 0.470 Volts 4.700 CLKGEN 1 clock 8 ADC 2369 ( 941) MUX 0.470 Volts 4.700 CLKGEN 1 clock 9 ADC 2389 ( 955) MUX 0.499 Volts 4.993 CLKGEN 1 clock 10 ADC 2456 ( 998) MUX 0.597 Volts 5.974 CLKGEN 1 clock 11 ADC 2456 ( 998) MUX 0.597 Volts 5.974 CLKGEN 1 clock 12 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 1 clock 13 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 1 clock 14 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 1 clock 15 ADC 2388 ( 954) MUX 0.498 Volts 4.978 CLKGEN 1 clock 16 ADC 2369 ( 941) MUX 0.470 Volts 4.700 CLKGEN 1 clock 17 ADC 2369 ( 941) MUX 0.470 Volts 4.700 CLKGEN 1 clock 18 ADC 2368 ( 940) MUX 0.469 Volts 4.685 CLKGEN 1 clock 19 ADC 2369 ( 941) MUX 0.470 Volts 4.700 CLKGEN 1 clock 20 ADC 2369 ( 941) MUX 0.470 Volts 4.700 CLKGEN 1 clock 21 ADC 2389 ( 955) MUX 0.499 Volts 4.993 CLKGEN 1 clock 22 ADC 2456 ( 998) MUX 0.597 Volts 5.974 CLKGEN 1 clock 23 ADC 2456 ( 998) MUX 0.597 Volts 5.974 Low-rail clock voltages for clock gen 2 CLKGEN 2 clock 0 ADC 1505 ( 5e1) MUX -0.795 Volts -7.950 CLKGEN 2 clock 1 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 2 clock 2 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 2 clock 3 ADC 1503 ( 5df) MUX -0.798 Volts -7.980 CLKGEN 2 clock 4 ADC 2046 ( 7fe) MUX -0.003 Volts -0.029 CLKGEN 2 clock 5 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 6 ADC 1506 ( 5e2) MUX -0.794 Volts -7.936 CLKGEN 2 clock 7 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 2 clock 8 ADC 1504 ( 5e0) MUX -0.796 Volts -7.965 CLKGEN 2 clock 9 ADC 1503 ( 5df) MUX -0.798 Volts -7.980 CLKGEN 2 clock 10 ADC 2047 ( 7ff) MUX -0.001 Volts -0.015 CLKGEN 2 clock 11 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 12 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 13 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 14 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 15 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 16 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 17 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 18 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 19 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 20 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 21 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 22 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 23 ADC 2048 ( 800) MUX 0.000 Volts 0.000 High-rail clock voltages for clock gen 2 CLKGEN 2 clock 0 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 1 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 2 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 3 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 4 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 5 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 6 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 7 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 8 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 9 ADC 2184 ( 888) MUX 0.199 Volts 1.991 CLKGEN 2 clock 10 ADC 2049 ( 801) MUX 0.001 Volts 0.015 CLKGEN 2 clock 11 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 12 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 13 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 14 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 15 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 16 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 17 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 18 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 19 ADC 2049 ( 801) MUX 0.001 Volts 0.015 CLKGEN 2 clock 20 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 21 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 22 ADC 2048 ( 800) MUX 0.000 Volts 0.000 CLKGEN 2 clock 23 ADC 2048 ( 800) MUX 0.000 Volts 0.000 Closed clock voltages are all within tolerance Now commanding shutter closed +5v readback raw ADC reading = 3419 volts= 5.013420 prior to clocking +15v readback raw ADC reading = 3486 voltage = 15.971211 prior to clocking -15v readback raw ADC reading = 617 voltage = -16.091669 prior to clocking +30v readback raw ADC reading = 3277 volts = 34.482758 prior to clocking Sending IDL command to TIMING board CCD clocks are now running in idling mode value = 0 = 0x0