Introduction
The XMET display is an informational widget providing current
meteorlogical information. XMET is a monitor program and offers no
controls.
To start XMET, right click on a blue background desktop
in your VNC session and from the popup menu select, depending on
the instrument:
K2 Telescope Status Menu → XMET - Weather
or:
Telescope GUIs → XMET - Weather
Stop XMET by using the Close option from the window
menu, accessible by clicking the button at the top left of the
window.
The observers version of XMET displays a graph of temperatures over
a selectable time period. Instantaneous values are displayed, as
well as the following values on the graph, which is updated every
minute:
- Dewpoint point, maximum (dark blue line)
- Primary mirror temperature, actual (green line)
- Secondary mirror temperature, actual (magenta line)
- Outside air temperature, minimum (cyan line)
- Dome temperature (red line)
There is a second line graph which displays the outside humidity
(cyan line) and dome humidity (red line) over a selectable period of
time. Both graphs are updated every minute.
XMET also displays the wind speed (mph) and direction. The wind
speed is shown as the instantaneous current value, the last minute
maximum and the last 30 minutes maximum. The wind direction is
represented in terms of the minimum, maximum and average value
within the last minute.
The stop-lite buttons on the right side of the display
warn if the dewpoint is approaching the specified temperature
(e.g., primary or secondary mirror temperatures or outside
temp). These calculations are based on instantaneous data and they
are color coded as follows:
-
Green: The dewpoint is 5 degrees
or more below the specified temperature.
-
Amber: The dewpoint is between 2
and 5 degrees below the specified temperature.
-
Red: The dewpoint is dangerously within 2
degrees below the specified tempeperature.
-
Pri-Dew: Difference between the primary mirror
temperature and dewpoint.
-
Sec-Dew: Difference between the secondary mirror
temperature and dewpoint.
-
Out-Temp: Difference between the outside temperature
and dewpoint.
There is also a precipitation indicator.
Warning messages
Warning text displayed in the red alarm box indicate the following:
-
NO MET DATA: means the met VME crate is possibly down or the
tcl scripts are not running (see Trouble-Shooting Section
below). When this happens, no data values or appropriate stop-lite
warnings will be displayed. You will notice that the linegraph
trends may 'spike downwards'. The data labels will also turn red.
-
STALE GE DATA; Check TCL Scripts: means that the data has
not changed for a while (see Trouble-Shooting Section below.)
-
INVALID GE DATA: dewpoint, outside air, and outside
humidity will turn red.
-
GE CLEANING: means that the GE unit is going through its
daily cleaning cycle. You will not see GE dewpoint, outside air,
and outside humidity since these values should be ignored during
these times. The appropriate line graphs will spike downward.
-
GE NEEDS SVC: means that the GE unit needs service. Follow the
instructions under the Electronics group homepage. You will not
see GE data values during this time. You may also see the data
spike downwards.
-
NO PRI MIRR DATA: means that the ACS is not in the correct
mode to output data. If the label for SECONDARY is red also,
this means that the AUX crate is down.