Background

The adctestAll command for the ADC checks software and settings and reports any anomalies.

Invocation

Execute the command from any LRIS account on the LRIS host computer via:
	adctestAll

Output

Normal output from the script is as follows:
Checking ADC daemons:
  Checking galil...................OK
  Checking pulizzi.................OK
Checking ADC devices:
  Checking galil...................OK (Galil controller)
  Checking pulizzi.................OK (Pulizzi AC power controller)
Checking ADC keyword libraries:
  Checking ADC.....................OK
  Checking DCS.....................OK
Checking ADC settings:
  Checking ADC status..............OK 
  Checking ADC limits..............OK 
  Checking remote control..........OK 
  Checking disconnected cables.....OK 
  Checking software lockout........OK 
  Checking calibration.............OK 
  Checking electronics rack temp...OK 
  Checking bypass switch setting...OK 
  Checking instrument ESTOP........OK 
  Checking facility ESTOP..........OK 
  Checking blue limit..............OK 
  Checking red limit...............OK 
  Checking control mode............OK 
  Checking ADC Prism Calcs.........OK 
  Checking AC outlet 1 (Galil).....OK 
  Checking AC outlet 2 (Interlocks)OK 
  Checking AC outlet 3 (Motor).....OK 
  Checking AC outlet 4 (Hub).......OK 
---------------------------------------------------------------
        All tested ADC systems tested appear functional.
---------------------------------------------------------------