NIRC/Guider Focus and Tilt Comparison

On 10 May 1998, Malign was run at various rotator position angles (ROTPPOSN) on both the NIRC detector and the guide camera. This provides a comparison of the focus and the secondary tilts between the two systems as a function of p.a.

From this, one could, for example, focus a star on the guider, offset the focus by the amount shown below, and provide a well-focused image to teh NIRC detector in situations in which one cannot focus directly on the NIRC detector. One such situation may be focusing at 3-5 microns, using the max12ur timing pattern. Note that elevation effects are not addressed here.


  NIRC Guider
ROTPPOSN Focus az tilt el tilt Focus az tilt el tilt
-180° -0.053 -1.674 1.062 -0.170 -21.730 -5.289
-150° -0.052 -1.063 1.632 -0.183 -18.960 -14.598
-120° -0.044 -1.264 0.721 -0.190 -5.558 -18.590
-90° -0.050 0.355 -1.205 -0.165 0.045 -19.213
-60° -0.050 -1.234 -2.387 -0.129 10.816 -18.553
-30° -0.040 -0.020 -1.673 -0.083 19.139 -9.359
-0.026 0.676 -0.510      
30° -0.015 0.287 -0.695 0.009 15.968 13.443
60° -0.014 4.440 -1.558 0.009 6.385 19.353
90° -0.001 2.172 -1.473 0.009 -6.371 18.423
120° 0.005 0.254 -1.857 -0.039 -10.333 14.819
150° -0.002 1.444 -3.184 -0.087 -16.879 10.138

ChartObject Guider Tilts

ChartObject NIRC Tilts

ChartObject Focus