Purpose
The NBC SDSU Gen-2 electronics contains a power control board
When power is applied to the electronics, only logic-level
power is turned on. +/- 16V and +34V are not turned on or
applied to the CCDs. This is a safety precaution.
This script sends a keyword to the NBC VME system which
executes a server routine to send the "PON" command to the
SDSU electronics.
It checks the value of the +34V power supply first and
skips the modify command if this power is already on.
Usage
nbc_pon
Arguments
none
Exit values
0 = normal completion
-1 = error