Characteristic Value Comments
Detector Hawaii-2 2048 x 2048
Minimum Exposure Time 2 s
Saturation 33,000 DN
Persistence >500 DN Occasionally seen at lower exposure levels
Bad Pixels
Typical Exposure Level 150 DN/s/wavelength channel 10th mag star, K-band, 20 mas scale, AO loops closed

SPEC: Measured Noise Characteristics

Characteristic Integration Time Mean Median Measurements
Dark Current 900 s 1.34 ± 0.60E-02 DN/s 1.34E-02 DN/s 112
Read Noise 900 s 12.81 ± 0.95 e- 12.52 e- 107

Long-term Monitoring Noise Plots

Dark Current Read Noise
    Notes:
  • Negative values are not understood
  • Values above 0.02 DN/s are thought to be shortly after a detector reset
    Notes:
  • Values above 14 e- are thought to be from images shortly after a detector reset

Noise Plots vs. Integration Time

Dark Current Read Noise
    Notes:
  • Dark current values in the shorter integration times appear significantly influenced by the reset anomaly of the Hawaii-2 detector.
  • The true dark current appears to be asymptotic value towards longer integration times.
    Notes:
  • Read noise values are uncorrected for the contribution by noise of the dark current.