| Video Input | Amplifier | Gain [e-/DN] |
Readnoise [e-] |
Dark Current [e-/pixel/hr] |
|---|---|---|---|---|
| 1 | 1A | 1.232±0.024 | 2.584±0.051 | -- |
| 2 | 1B | 1.178±0.015 | 2.470±0.032 | 4.19 |
| 3 | 2A | 1.698±0.037 | 1.781±0.038 | -- |
| 4 | 2B | 1.171±0.015 | 2.455±0.031 | 3.46 |
| 5 | 3A | 1.190±0.009 | 2.495±0.019 | -- |
| 6 | 3B | 1.317±0.027 | 2.761±0.057 | 4.03 |
| 7 | 4A | 1.251±0.015 | 3.933±0.049 | -- |
| 8 | 4B | 1.216±0.018 | 2.549±0.039 | 3.80 |
| Video Input | Amplifier | Gain [e-/DN] |
Readnoise [e-] |
Dark Current [e-/pixel/hr] |
|---|---|---|---|---|
| 2 | 1B | 1.179±0.020 | 2.472±0.041 | 3.30 |
| 4 | 2B | 1.174±0.014 | 2.461±0.029 | 3.60 |
| 6 | 3B | 1.312±0.017 | 2.752±0.035 | 3.50 |
| 8 | 4B | 1.212±0.015 | 2.541±0.032 | 3.70 |
| 10 | 5B | 1.213±0.018 | 2.542±0.038 | 2.70 |
| 12 | 6B | 1.198±0.019 | 2.512±0.039 | 3.80 |
| 14 | 7B | 1.204±0.019 | 2.524±0.039 | 3.30 |
| 16 | 8B | 1.249±0.016 | 2.619±0.033 | 3.70 |
| Gain [e-/DN] |
Readnoise [e-] |
|---|---|
| 16.266± 0.011 | 53.561± 1.991 |