Slit Viewing Camera

The instrument also has a near-IR slit viewing camera.


Camera schematics from Wilson et al. 2004. The actual filter is a K'.

Filter

The slit viewing camera has a K' filter.

type central wavelenth
[micron]
band wavelength
[micron]
temperature
[K]
Notes
Mauna Kea K' 2.123 0.337 77 Barr Associates


Filter data.

Detector

The detector is a Rockwell Scientific Hawaii-1 array (engineering grade).

array size pixel size
[micron]
plate scale
[arcsec per pixel]
Field of View
[arcmin]
Notes
1024x1024 18.5 0.12 2x2 engineering grade

Detector parameters

gain
[e-/ADU]
readnoise
[CDS in electrons]
linearity
(<10% in ADU)
Full well
[electrons]
minimum integration time
[sec]
4.98 26 35000 160000 1.2


Slit viewing camera image. It shows the slit (on the middle left).


Go To:
Instruments Home Page       Keck Home Page       Observer's Reference Shelf


Last modified: 06/14/2017 01:28
Send questions or comments to:NIRES Support

The information on this page is the property of the W. M. Keck Observatory. The contents of this page or any part thereof shall not be copied or otherwise reproduced or transferred to other documents or used or disclosed to others for any purpose other than observing support at the W. M. Keck Observatory and the subsequent analysis and publication of scientific data obtained from observations conducted at the W. M. Keck Observatory. All rights reserved. © W. M. Keck Observatory.