Log Directory: /kroot/data/mosfire/logs/housekeeping/
| Keyword Service | Function | Filename | Format | Number of | Columns | ||
|---|---|---|---|---|---|---|---|
| Columns | Name/Kwd | Units | Description | ||||
| MTCS | Dewar Temp Control | yymmdd_mtcs.log | CSV | 15 | Date | yy/mm/dd | HST Date |
| Time | hh:mm:ss | HST Time | |||||
| curtmp1 | K | Detector Mount | |||||
| curtmp2 | K | +X Bulkhead A (CCR #1 side) | |||||
| tmp3 | K | Field Lens Cell | |||||
| tmp4 | K | FCS Mount | |||||
| tmp5 | K | Camera Barrel | |||||
| tmp6 | K | Collimator Barrel | |||||
| tmp7 | K | -X Main Spreader Bar (CCR #2 side) | |||||
| tmp8 | K | +X Main Spreader Bar (CCR #1 side) | |||||
| tmp9 | K | Cold Head #3 End of DH Thermal Link | |||||
| tmp10 | K | Cold Head #3 Low Temp Stage | |||||
| output1 | % | Channel A Output | |||||
| output2 | % | Channel B Output | |||||
| comment | n/a | Usually blank | |||||
| Example: | 18/08/28,15:23:50, 77.000, 120.003, 120.463, 120.642, 120.861, 120.484, 111.904, 110.706, 61.379, 10.832, 10.5, 51.2, | ||||||
| Keyword Service | Function | Filename | Format | Number of | Columns | ||
| Columns | Name/Kwd | Units | Description | ||||
| MDHS | Housekeeping | yymmdd_mdhs.log | CSV | 16 | Date | yy/mm/dd | HST Date |
| Time | hh:mm:ss | HST Time | |||||
| tmp1 | degC | Electronics Cabinet Air Return Temperature | |||||
| tmp2 | degC | Electronics Cabinet (between) CSU Chassis Temp | |||||
| tmp3 | degC | Electronics Cabinet "right side middle back" Temp | |||||
| tmp4 | n/a | Not Used (shows "Error") | |||||
| tmp5 | degC | Dewar Window Temp | |||||
| volt1 | V | Dewar Window Heater Voltage | |||||
| cur1 | A | Dewar Window Heater Current | |||||
| volt2 | V | Detector Heater Voltage | |||||
| cur2 | A | Detector Heater Current | |||||
| ebaydoor | string | Electronics Cabinet Door [Open|Closed] | |||||
| glysup | string | Glycol Flow [Flow|No flow] | |||||
| glyret | string | Glycol Flow [Normal|Overflow] | |||||
| hpsupply | string | Power Supply [Normal|Crowbarred] | |||||
| comment | n/a | Usually blank | |||||
| Example: | 18/08/28,15:23:09, 2.97, 1.84, 4.83, Error, 3.05, 20.19, 0.788, 51.32, 0.811, Closed, Flow, Normal, Normal, | ||||||
| Keyword Service | Function | Filename | Format | Number of | Columns | ||
| Columns | Name/Kwd | Units | Description | ||||
| MPAS | Dewar Pressure | yymmdd_mpas.log | CSV | 7 | Date | yy/mm/dd | HST Date |
| Time | hh:mm:ss | HST Time | |||||
| g1str | Torr | TC1 - low vacuum gauge | |||||
| g2str | n/a | TC2 - unused (reads -30 Torr) | |||||
| g3str | Torr | ION1 - high vacuum gauge #1 | |||||
| g4str | Torr | ION2 - high vacuum gauge #2 | |||||
| comment | n/a | Usually blank | |||||
| Example: | 18/08/28,15:24:42, 1.000e-04, -3.000e+01, 1.586e-08, 6.971e-08, | ||||||