Follow these steps at the end of the night to turn LWS off:
- Stop data acquisition.
In the Xpose/LWS window, click
on Abort to halt any current
exposure.
- Insert ZnSe window.
In the XLWS window, click on Window IN and SET to re-insert the protective
cover over the dewar window.
- Insert 11.7 µm filter.
In the filter area of the XLWS
window, click the 11.7 µm
filter button and then click SET
to insert this filter.
- Turn off detector.
In the XLWS window, click on the
power OFF and heater OFF buttons to turn off the
level power and detector heater, respectively.
- Stop instrument software.
- From the window manager's main menu, select LWS Control and then LWS Shutdown. This will shut down
most of the LWS windows. Please verify that the ZnSe dewar
window is returned to its IN position.
- Open the iconified LWS Start window and
press CTRL-C in order to interrupt other
windows.
- Stop Quickview software.
- Click the File droplist in
the Quickview display and release on Quit.
- In the Quickview console window, type
exit to end the IDL session and close the
window.
- Stop remaining software.
Kill any existing FACSUM and guider
eavesdrop windows by clicking and holding the right mouse button in the window's
titlebar and releasing on the QUIT
item.
- Log out.
From the window manager's pulldown menu, select the
Exit... item to end your Xwindows session and log
out of the workstation.