It is important to determine the correct instrument focus before
performing the telescope focus procedures (a.k.a. MIRA ) to avoid incorrect results.
The instrument focus is determined by measuring the FWHMs of comparison arc lines along the spectrum.
Polarimeter focus data are obtained with arc lamp exposures on both ordinary and extraordinary beams.
To facilitate rapid and efficient measurements of instrument focus, Keck provides
software to configure the instrument, to gather focus images and to analyze them.
See below for setup and procedures.
The default instrument setup for taking polarimeter focus data is spectroscopic mode on both red and blue arms with
either the 1.0 or 1.5 arcsec wide polarimeter long slits in the beam:
- Configure the instrument for focusing.
trapdoor | closed |
slitmaks | pol_1.0 or pol_1.5
|
dichroic | as desired |
grism | as desired |
grating | as desired |
wavelen | as desired |
red filter | clear |
blue filter | clear |
blue focus start | -2450 |
red focus start | -0.10 |
blue/red exp. time | 1 sec |
- Run the spol.window script to window red CCD for polarimeter readout.
- Turn on arc lamps:
- Red side: Switch on the Ne, Ar lamps.
- Blue side: Switch on the Hg, Cd and Zn lamps. Wait ~5 min for lamps to warm up.
- Start the focus loop for spectroscopy
From the CMD...
menu select Focus loop Spectroscopy
This will bring up a gui with default
parameters. Click the GO button to
initiate the loop.
When the first images are displayed check that the signal level of the "slit" is
well above the background (at least 4000 counts above).
Examples of focus
data for the red and blue sides are shown below.
Blue side image of long slit used to focus the instrument.
|
Red side image of long slit used to focus the instrument.
|
- The script will take the requested number of images and save them in the default
data directory as bfocXXXX.fits and rfocXXXX.fits, where XXXX is the current frame
number for blue and red.
- Analyze the focus images
- Launch IRAF: Right-click on the background pane and
select from the pulldown menu:
IRAF...cl + ds9
- Load keck.lris package
- "cdata" to change to the data directory
- run lspecfocus. A prompt will
request either the "last 7 red" or "last 7 blue"
- use your cursor to mark the regions
encompasing the spectrum.
Mark the region
- The routine will next plot the FWHM of the lines and the best
fits at each of the focus positions and fit a quadratic to the data
Best fit.
- Enter the best focus value into the XLRIS focus setting.
- In a punaluu window type:
focus_table | lpr -Plw4
The focus table will be printed on the laser printer located in the same room, close to the entrance.
Just typing focus_table in the punaluu xterm will print the table on screen. See below for example output
from focus_table
Do not change the instrument configuraiton before moving the instrument.
The focus_table reads the current setup before calculating the focus table. If the instrument
configuration is changed after focusing, the focus_table script will report incorrect values.
focus_table Example output
>focus_table
R E D S I D E F O C U S
Filter Imaging Longslit
+ multislit spectroscopy
+ superlong
==============================================
clear 2342 2301
B 2237 2196
V 2442 2401
R 2282 2241
Rs 2307 2266
I 2288 2247
GG495 2336 2295
OG570 2230 2189
RG850 2334 2293
Dichroic add offset:
mwmwmwmwmwmwmwmwmwmwmw
clear 0
mirror 0
460 -7
500 11
560 16
680 -9
B L U E S I D E F O C U S
Filter Imaging Longslit
+ multislit spectroscopy
+ superlong
==============================================
clear -2834 -2799
U -2809 -2774
B -2854 -2819
G -2834 -2799
V -2833 -2798
NB4170 -2859 -2824
There is no dichroic offset for the blue side