LRIS
Red CCD Characteristics

Note: This is for a previous generation detector on LRIS, not the current detector.

In June 2004 the clock driver board failed. Some capacitors and the ADC were replaced.

In the period 2002-May-9 to 2003-July-28 the detector electronics was not working properly and, as a result, strong blooming was generated by pixels at the ~34,000 ADUs level or higher. 

Board Service
Dates
Measure
Date
Gain*
[e-/ADU]
Noise*
[e-/pixel]
Bias*
[ADU]
Linearity 

[up to ... ADU]

Left Right Left Right Left Right Left Right
CCD-2 Leach Gen 1 - After repair 2003-June-10 2003-June-13 1.98 2.17 6.1 6.3 ~752 ~850 ~60,000 ~62,000
CCD-2 Leach 1 system 2003-Aug-07 2003-Aug-7 1.98 2.21 6.2 8.6 ~735 ~835    
CCD-2 Leach 1 system + temp electronics 2003-July-31
-
2003-July-29
2003-July-28 1.46 1.36 6.2 6.8 - -    
CCD-2 Leach 1 system 1996-Jun-24
-
2002-May-8
1996-Jun-26 1.97 2.10 6.3 6.6 ~990 ~1080    
Combo CCD-1 Leach system 1995-Dec-7
-
1996-Jun-24
1995-Dec-8 1.83 2.16 5.34 5.53 ~850 ~920    
Spare Leach system 1995-Oct-17
-
1995-Nov-15
1995-Oct-20 1.38 1.43 9.9 10.1 ~727 ~793    
Primary Leach system 1995-Jul-20
-
1995-Oct-17
1995-Aug-28 2.19 2.19 6.65 6.16 ~940 ~968    

*All values refer to low gain mode.


Go to: LRIS Home Page - Instruments Home Page - Keck Home Page

LRIS Master
Last modified: Aug 7, 2002