![]() |
ESI Diagnostic Checklist | ![]() |
Hi folks!
In case you have problems bringing up ESI & accessing keywords, here are some possible solutions Following is a summary of what was done today to get access to the relevant ESI keywords (IONPUMP, etc).
Thanks!
Liz
P.S. You can also check out these web pages :
The ESI Dispatchers and Galil Controllers
You bring up Dashboard (ESI Display) from the 'esicon' menu and it seems to hang. The border comes up, but no display 'face,' and the console window says it can't read ESI keywords :
INIT: Reading Layout File /kroot/data/esi/esi.dbd esi.tlib ERROR reading bin KW esi DWFILNAM: Can't read ESI dispatcher status: mread select timeout. ERROR reading bin KW esi SLMSKNAM: Can't read ESI dispatcher status: mread select timeout.
If you can, your dashboard should now come alive. Otherwise, kill it ('kill_em obsr') and restart it.
You went thru the above step (1) and you still can't read keywords.
Log in as esieng to koki:
If you don't get a response, power-cycle the Lantronix. If it was not power-cycled after the instrument was power-cycled, please power-cycle it now.
Go To: | ||||
Instruments Home Page | Keck Home Page | Observer's Reference Shelf |
ESI Master
7 February 2000