Name
cycle_shutter -- perform an open/close sequence on the DEIMOS CCD shutter
Name
cycle_shutter [n]
Type
C-shell script
Purpose
perform a close-open-close sequence on the DEIMOS CCD
shutter. This action is often useful in freeing a stuck
shutter, thus recovering from the common error condition in
which the datataking system refuses to start an exposare.
Invocation
cycle_shutter [n]
Inputs
n = number of times to cycle shutter. IF n<1 we cycle until
the shutter operates successfully or a max of n tries.
Exit status
0 = normal completion
1 = error
See Also
The shell script cycle_shutter
DEIMOS script index
Go to:
DEIMOS Home Page -
Instruments Home Page -
Keck Home Page
Last modified: Wed Apr 13 09:36:25 2005