Background

DEIMOS contains a Photometrics PXL-series TV guider system similar to those used on ESI, NIRSPEC, and ACAM. Accessing the DEIMOS guider is a laborious and time-consuming process requiring major disassembly of the instrument's front end. This document describes the steps taken during an emergency effort to service the camera in August 2002.

Resource Summary

Procedure

  1. Position telescope. Align telescope so that the tube elevation is greater than 45° and azimuth position aligns DEIMOS (NRT1 or NRT2) with RT4 on the Nas Deck. Lock out and tag out telescope for az or el moves.
  2. Position DEIMOS. Drive DEIMOS back far enough to gain ample access to front of snout:
  3. Rotate DEIMOS.

    Figure 1. View of snout from front

  4. Disassemble front end. Remove components on the DEIMOS snout to gain access to the camera region.
  5. Protect Optics The window under the hatch has a very delicate optical coating which cannot be touched. Protect this coating as follows:

    Figure 5. Protective cloth applied to window

  6. Disconnect Air & Glycol.
  7. Remove camera.