Single segment guiding is a way of obtaining slit spectroscopy without using either on-slit guiding or using a second star that lies in the same guider field of view as the science target.
In order to accomplish this, we use a unique character of a segmented primary mirror; the ability to take one of the segments out of the stacked image, thus creating a “satellite” image of the science target. This satellite image can then be used as a guide star, with some modifications to normal procedure. This seems to be most relevant to HIRES, where the bright science targets, the small guider field of view, and the ever present slit (!) can combine to prevent good observations of some targets. Observers have been known to skip observing some of their targets that disappear almost entirely down the slit, leaving nothing to guide on.
There may be an additional benefit. Even if enough light spills out of the slit to allow on-slit guiding, off-slit guiding on a satellite image may produce higher quality data. Tests still need to be designed and carried out to determine how much benefit, if any, this provides.
On the ACS CLI, type:
tilt segment 1 3 0
The parameters are: segment number, segment tilt in arcsec in the X direction (elevation), and segment tilt in arcsec in the Y direction (azimuth). Note that the segment tilt is half the distance on sky that the segment’s image will move. So in the example above we tilt segment 1 by 3 arcsec in the azimuth direction, moving it’s image 6 arcsec on the sky. (In many cases the slit will be vertical, so a tilt in azimuth will move the segment image perpendicular to the slit.) Note that the tilt segment command is absolute, so typing tilt segment 1 3 0 a second time will not move the segment farther from the main image.
Also note that when a segment is tilted the ACS sensor gains will change, leading to some degradation in control. It is expected that this will be small. However, some restraint in how far you throw the segment should be considered.
Finally, note that the ACS will take about 30 seconds to settle. You may seem the stack fall apart momentarily, so beware of performing this step during an exposure.
Table 1. Differential guiding parameters vs. rotator mode and instrument for single-segment guiding.
Rotator Mode |
Instrument |
Differential guiding parameters |
|
---|---|---|---|
sEL |
sPARANG |
||
VERTICAL |
all |
Turn differential guiding off
|
|
POSITION ANGLE |
all |
0 |
–1 |
STATIONARY
|
NIRC, LWS, ESI, LRIS (any Cassegrain instrument) |
Turn differential guiding off
|
|
DEIMOS, HIRES, NIRC-2, NIRSPEC, OSIRIS (any Nasmyth or bent Cassegrain position) |
–1 |
0 |
While watching the stars and segment image on the guider, the main thing you are concerned about is that the star’s primary image stay fixed in the middle of the slit. If this happens, you are guiding properly. There are, however, some useful cues that can indicate problems and their solution: