Purpose
This checklist leads you through the set procedure for the
DEIMOS Flexure Compensation System (FCS).
Preparing DEIMOS for FCS setup
- Configure spectrograph. Configure the instrument
with the desired grating, tilt wavelength, and filter for
acquiring spectra.
- Close hatch. Make certain that the
remotely-controlled hatch at the front of DEIMOS is closed.
- Extinguish lamps. Make certain that all of the
internal calibration lamps are turned off.
- Insert slitmask. To reduce the amount of stray
light that reaches the dewar, insert a slitmask. Any slitmask
will do, even the focus holes mask or the longslit.
- Idle tracking. Make sure that the FCS tracking
loop is in the idle state. Check the FCS
MODE status on the left side of the main panel of the
DEIMOS FCS GUI; it should
indicate Off and the FCS STATUS display on the right side of
this panel should indicate idle.
If not, then click on the Stop
button, and wait until the FCS STATUS displays idle.
- Enable rotator. Bring up the Rotation sub-panel from the the
main DEIMOS Dashboard GUI. If
the rotation status indicates that rotation is
locked, press the UNLOCK button.
Taking the Reference Frame
The first set of instructions below pertain to the normal case
of observing a slitmask without regard to previous observations.
If you are re-observing a slitmask that was observed on a
previous night of your run, then follow the
matching procedure below. If you are
re-observing a slitmask that was observed on a previous run,
then follow the
archival matching
procedure below.
Normal (Non-Matching) Case
- Launch Setup panel. On the main panel of the DEIMOS FCS GUI, click the Setup button. This will bring up the FCS Setup sub-panel.
- Reset exposure time. On the right-hand side of the
Setup subpanel, click the Exposure
Time button and select 1.
- Select mode. Near the bottom of the FCS SETUP
sub-panel, select New for
Choose reference frame type.
- Home FCS Stages. On the Setup Subpanel, click the
button in the lower left corner reading Prepare for new reference frame. Wait
for the message box at the bottom of that subpanel to show
fcszero new successful. You are now ready to take reference images
- Acquire reference image. When the FCS system is
ready, click on Take new reference
frame and select Spectrographic when prompted. The system
will acquire an FCS exposure.
- Adjust FIGDISP. The newly-acquired reference image
should appear on the FIGDISP image display on your screen. For
the first image read out, you should do the following on
the tk control window:
- Click the NORMAL and
CENTER buttons to restore
normal zoom and pan factors.
- Click HISTEQ to select the
histogram-equalized colormap.
- Click IN to zoom in until
the window is filled.
- Click the green Stats Box
button to put the mouse into statistics mode.
- Measure peak flux. Draw a statistics box around the
2 spectra on the left half of the FIGDISP display by clicking
and dragging with the left mouse button. In the resulting popup
window, note the max value. Draw another statistics
box around the 2 spectra on the right-hand side and again note
the max value.
- Adjust exposure time. If the peak flux in each
statistics box is at least 2000, the reference frame is
acceptable; proceed to the next step. If not, then select a
longer expoxure time by clicking the Exposure Time button on the Setup subpanel
and choosing a longer exposure time. Click on Take new reference frame and select Spectrographic when prompted. The system
will acquire an new FCS exposure. Repeat the previous step to
measure the peak flux and continue to lengthen the exposure time
until both sides have a peak flux over 2000.
- Verify Tracking. Click Start on the main FCS GUI to enable FCS. The FCS
Status should report OK immediately. Within a minute, FCS
should lock on with messages of Tracking and On Target, as
shown here. If FCS Status reports
Lockout, ensure the shutter is open. If FCS Status still does
not report OK, please contact your support astronomer.
- Repeat as needed. Repeat all steps in this section
for each unique combination of grating, central wavelength, and
filter you plan to use during the night.
Matching Against a Previous Night
Follow this procedure if you are re-observing a slitmask that
was observed on a previous night of your run. Note: if you want
to match against observations made on a previous observing run,
please see the
procedure below.
- Launch Setup panel. On the main panel of the DEIMOS FCS GUI, click the Setup button. This will bring up the FCS Setup subpanel.
- Verify match. On the FCS Setup subpanel, ensure that
the Current Reference File, Current Image File, and CurrentLog File are all defined. If they are
not, then either the reference file from the previous night
needs to be copied into the FCS directory, or the instrument is
not configured with exactly the right grating, wavelength, and
filter to match the previous setup.
- Select mode. Near the bottom of the Setup
sub-panel, select Matching for
Choose reference frame type.
- Home FCS Stages. On the Setup subpanel, click the
button in the lower left corner reading Prepare for matching reference frame.
Wait for the message box at the bottom of that subpanel to show
fcszero match successful. You are now ready to take reference images
- Acquire reference image. When the FCS system is
ready, click on Take matching reference
frame and select Spectrographic when prompted. The system
will take several minutes to reset the focus to the value
employed in the old FCS setup, run FCS to achieve FCS lock
against the old FCS reference image, restore the focus value,
and acquire a new FCS reference exposure. Your indication that
the procedure is done will be when the Current
Image File shown on the FCS
Setup panel changes.
- Adjust FIGDISP. The newly-acquired reference image
should appear on the FIGDISP image display on your screen. For
the first image read out, you should do the following on
the tk control window:
- Click the NORMAL and
CENTER buttons to restore
normal zoom and pan factors.
- Click HISTEQ to select the
histogram-equalized colormap.
- Click IN to zoom in until
the window is filled.
- Click the green Stats Box
button to put the mouse into statistics mode.
- Measure peak flux. Draw a statistics box around the
2 spectra on the left half of the FIGDISP display by clicking
and dragging with the left mouse button. In the resulting popup
window, note the max value. Draw
another statistics box around the 2 spectra on the right-hand
side and again note the max value.
- Adjust exposure time. If the peak flux in each
statistics box is at least 2000, the reference frame is
acceptable; proceed to the next step. If not, then select a
longer expoxure time by clicking the Exposure Time button on the Setup subpanel
and choosing a longer exposure time. Click on Take new reference frame and select Spectrographic when prompted. The system
will acquire an new FCS exposure. Repeat the previous step to
measure the peak flux and continue to lengthen the exposure time
until both sides have a peak flux over 2000.
- Verify Tracking. Click Start on the main FCS GUI to enable FCS. The FCS
Status should report "OK" immediately. Within a minute, FCS
should lock on with messages of "Tracking" and "On Target", as
shown here. If FCS Status reports
"Lockout", ensure the shutter is open. If FCS Status still does
not report "OK", please contact your support astronomer.
- Repeat as needed. Repeat all steps in this section
for each unique combination of grating, central wavelength, and
filter you plan to use during the night.
Matching Against a Previous Run
Follow these steps to set up FCS to match against a reference
image acquired on a previous run.
- Determine the UT date of the original observation.
- Locate the corresponding subdirectory for that UT date in
/home/dmoseng/fcs_reference (on
deimosserver). For example, FCS
reference files for 2009-Oct-01 may be found in the directory
/home/dmoseng/fcs_reference/2009oct01.
- Copy the old reference files into the new FCS directory by
issuing these commands on deimosserver:
cdfcs
retrieve_fcsref date
- Continue with the procedure for matching described above.
Note: If DEIMOS is already configured with the correct
grating, wavelength, and filter, then you will need to change
away from this configuration (e.g., by tilting the grating 100
units and then coming back) in order for the FCS system to
recognize that new FCS reference files exist in the FCS
directory.
See Also
Go to:
DEIMOS Home Page -
Instruments Home Page -
Keck Home Page
Last modified: Wed Oct 21 22:07:53 HST 2009