Purpose
This checklist describes the procedure for aligning
multiobject slitmasks on DEIMOS, assuming you are:
- using slitmasks which have alignment boxes
- using the xbox software for slitmask alignment
Procedure
- Focus telescope.
- Tell the OA the name of your target as
listed in your starlist file and also the position
angle for the instrument rotator.
- To save time, configure the instrument to run
Mira while the telescope is slewing:
- Set hatch to OPEN
- Set slitmask to None
- Set grating to:
- slider 3 or 4 if you plan to
align using the grating for imaging (remember to
set to zeroth order imaging position by pressing
the GO ZERO button), or
- to mirror if you plan to align using
the mirror for imaging
- Set filter to R
- Ask to OA to run Mira in the neighborhood
of your target.
- Wait about 10 minutes for the OA to acquire an
appropriate star and run Mira.
- Acquire target.
When the OA tells you that your target is on DEIMOS, check
the following information on FACSUM to verify that your
position is correct:
- Target name and coordinates
- Pointing origin (should be DEIMOS)
- Position angle (skypa on FACSUM). Note: also verify
that the rotator is not rotating into a limit by checking
the rotator screen on the dashboard.
- Complete rough alignment.
The goal of this step is to put the alignment stars into
their boxes. Ideally it will be completed the easy way
by positioning a star to a given position on the guider;
failing that, it will be completed the hard way by taking an
image on the science detector.
- The Easy Way.
One of the pieces of information output by the
Lick-supplied slitmask design software is the desired
coordinate at which you should place a fiducial star in
order to get the slitmask coarsely aligned. To do so:
- Identify the fiducial star on the guider image.
- Place the cursor on the star and note the
current (x,y) coordinates.
- At a polo prompt, use the
gmov command to perform a telescope
offset from the current coordinates (x1,y1)
to the desired coordinates (x2,y2):
gmov x1 y1 x2 y2
- The Hard Way.
If the Easy Way fails to work, you can try this
alternative method:
- Configure the instrument for imaging:
- select slider 3 and set to zeroth order
imaging position, or insert mirror
- set slitmask to None
- set CCD readout mode to Direct
- set exposure time to 20 sec
- Acquire first exposure.
- Insert the appropriate slitmask.
- Acquire second exposure.
- Using FIGDISP or IRAF, identify one alignment star
on the first image and measure the pixel coordinates.
Call these x1 and y1.
- On the second image, identify the corresponding
alignment box for the alignment star and note the
coordinates of the box center. Call these
x2 and y2.
- Move the telescope by typing the following command
into a polo window:
mov x1 y1 x2 y2
The alignment stars should now be in their boxes;
proceed with fine alignment.
- Complete fine alignment.
In this step, you will refine the positions of the stars within
their respective alignment boxes to achieve optimal
centering of the slitmask.
- Have the OA select a guide star and begin guiding.
Ensure that the guide star selected is not too close to the
edge of the guider field, since small offsets may still be
required.
- Start a 20-40 sec direct exposure through the
slitmask.
- In your IRAF window, type do_xbox anytime
before the image has been written to disk. The script
will wait for the current image to be written, will
determine the name of the image, and then will beep to
indicate that it's time for you to wake up and run
xbox.
- To run xbox on your image:
- Set up for spectroscopy.
- Reconfigure the instrument for spectroscopy:
- Set grating to the appropriate central wavelength
(if using zeroth order imaging you can use the GO BACK button to return to
original setting).
- Set filter to the desired position for order blocking.
- Set dewar focus as appropriate for the
selected filter.
- Set the detector parameters appropriately:
- Set ReadOut to Spectral.
- Set ObsType to Object.
- Set ExpTime as desired for spectroscopy.
- Wait for FCS to settle.
Do not start your exposure until the FCS GUI shows the FCS
status as locked, or it may be smeared. Hint:
read about the best way to
start a science exposure with DEIMOS.
- Expose!
See Also
Go to:
DEIMOS Home Page -
Instruments Home Page -
Keck Home Page
Last modified: Sun Nov 23 23:40:22 HST 2008