AO Calibrations Check

Daily AO calibration confirmation (affirmation).
  1. Start all LGS software by typing lgsao_start_vnc in a k2aoserver xterm. The xterm should be positioned in the upper left virtual window of the control-0 desktop when the command is issued.
  2. Start CalTools using the desktop menu: K2 AO tools -->Calibration Tools
  3. Kill the AOACQ tool.
  4. Run the Calibration Setup. From Keck II Calib gui select Calibration setup
  5. Run the light source gui:
  6. Restart the AOACQ GUI in simulator mode:
  7. in a kealoha xterm (or local xterm) run sounds kealoha so as not to annoy the summit staff with the sounds.
  8. In the AOACQ GUI do the following:
  9. start and setup the instrument.
  10. Using the AOACQ gui: click acquire star
  11. On the MAORI, set the gains to 0.1.
  12. check image on the instrument. For NIRC2 the PSF is typically round with FWHM of 45 mas. Elongated PSF \ or PSF with FWHM > 50 mas is considered poor. .
  13. If you need to check other pointings, repeat 7-11 for other pointings
  14. Run the light source gui:
  15. Using the AOACQ Gui:
  16. On the MAORI, set the gains to 0.05 (TT) and 0.2 (DM).
  17. Acquire an image on the instrument of choice and retest the image quality.
  18. On the Low-Bandwidth manager gui
  19. Using the MAORI gui:
  20. Check that LBWFS cog file is current
  21. Confirm that LBWFS does not sense focus or aberration (defoc < 0.2, and the graph looks flat). You may need to start looping.
  22. Check for 5x5 mode (the default was 20x20).