K2 NIRSPEC Procedures 28 December 2000 Edition
114-36-11
If Cold Cycle Refrigeration CCR) system is running, i.e. NIRSPEC
is COLD, then follow ALL steps in the procedure.
If NIRSPEC is WARM and the CCRs are off, then
you may omit steps in blue
lettering.
WARNING:
Be expeditious while the CCRs are off: Turning off
the NIRSPEC CCRs provides approximately 10 min before the dewar vacuum
pressure reaches an unacceptably high value. At that point, either the
CCRs need to be powered on or a vacuum pump needs to be pumping on the
dewar. Otherwise NIRSPEC will lose vacuum completely and will need to be
warmed to ambient, pumped, and then cooled back to operating temperature,
a 15 day process.
TOOLS
-
Stopwatch
-
CCR wrench kit from the NIRSPEC shelf in the instrument lab.
-
1 20 ft. extension cord.
-
1 power strip
-
1 TOSS 50 vacuum pumping station w/:
-
1 KF25 - KF40 adaptor
-
2 lengths of KF40 vacuum tubing
-
2 KF40 O-rings and clamps
-
Tools to remove the AO threshold.
-
1 Wire cutter.
PERSONNEL
-
1 person to operate telescope controls, switch cables in the computer room,
start/stop CCR compressors.
-
2 people at Nas Deck level to move the NIRSPEC handler and vacuum pump
and disconnect/reconnect cables.
-
1 person at Nas Deck level with stopwatch to keep track of time and monitor
vacuum pressures and read procedures.
TIME
-
Total Procedure should take about 2:20.
-K2 NIRSPEC Left Nasmyth AO Installation-
COMPUTER ROOM
-
Log on to Sun workstation "kua" as user "nspeceng".
-
From the workspace pulldown menu, choose NIRSPEC Control Menu->Stop
NIRSPEC Server.
TELESCOPE
-
Move telescope in elevation above 45 degrees to clear Nas Deck elevation
limit.
-
Move telescope in azimuth to 10.8 degrees to roughly align Nas Deck handler
position RT4 with AO enclosure door.
AO ENCLOSURE
-
Remove AO door clip and open AO door from the outside.
-
Remove AO door threshold. Stow threshold on Nas Deck. Picture
-
NOTE: the following steps apply only if the Dual Star Module (DSM) is
not present in the AO enclosure. If the DSM is present, move it out
and onto the "NIRSPEC storage rails" on the Right Nasmyth platform.
-
Remove rack and rail floor panels [labels to be determined by drew] in
order [also TBD by drew] inside AO enclosure. Stow panels on Nas Deck.
Picture
-
Install instrument insulated floor panels between rack and rail channels.
Picture
CAUTION: Ensure panels are down flat on the
flashing otherwise damage to the panels and/or instrument cart could occur.
-
Remove instrument interface access hole floor panel. Picture
-
Pull and fully extend the stainless steel braided CCR supply and return
lines [AO-CCR] out through the floor opening.
-
Locate and install the CCR Power patch cable,
AOCCRAC, to J??? on the AO/NIRSPEC Instrument
connect Panel (acronym? label?) and extend
the cable out through the floor opening.
-
Replace the grating with CCR lines and CCR power patch cable passing through
the slot between the grating and rail channel. Picture
-
Ensure the rack and rail, kinematic mount and fairlead areas are all clear
of obstacles.
-
Close AO door and clip shut.
TELESCOPE
-
If NIRSPEC is in the operational position, move telescope in azimuth to
193.7 degrees to line up Nas Deck handler position RT4 with the Right Nas
Platform NIRSPEC focus.
RIGHT NASMYTH PLATFORM
-
Verify that the rack and rails are clear of objects and debris between
NIRSPEC and the end of RT4
-
TEMPORARY: IN EFFECT 1 May 2001 Verify that the air supply
hose from the telescope is disconnected from the
regulator. The regulator is on or near the NIRSPEC glycol interface
panel, and has small-gauge plastic hose coming out of it and going
into the calibration unit. This hose carries the airflow for the
dewar window purge.
-
Open front right insulation panel on NIRSPEC and turn off main power breaker
on Pulizzi unit. Picture
-
Remove grating cover near boom and close instrument cooling supply ball
valve (valve handle points away from the cable boom), then close return
ball valve (valve handle points towards the cable boom). Picture
-
Connect transfer cart power cord from RT4 to NIRSPEC cart. Picture
-
Ensure that the cable boom tether is attached to the tension switch and
the tension switch is connected.
-
Unlock the cable boom deadbolt, located at the base of the boom.
-
Jack NIRSPEC up off the kinematic mounts by pressing and holding the "up"
button on the side of the handler until the jack screws stop automatically.
-
Release the cable ties to create slack in the cables.
-
Remove platform and deck safety chains, then drive NIRSPEC on its cart
to the RT4 park position on the Nas Deck. Carefully shepherd the boom cables
and boom rotation.
-
Re-install safety chains on Nasmyth platform and deck.
-
Cover Right Nas Platform kinematic mount cover plates with the steel plates
stored near the motor at the rear of the NIRSPEC handler.
-
NOTE: DO NOT REMOVE CCR HOSES OR CCR POWER CABLE
YET. Disconnect only the following boom cables:
-
Disconnect P2 & P3 (transputer fiber optics)
from Nirspec Connect Panel [NCP], J2 & J1.
-
Disconnect P4 & P5 (PXL Camera fiber optics)
from NCP, J4 & J3.
-
Disconnect NIRSPEC AC clean and commercial power cable.
-
Disconnect the Glycol supply and return lines from
the NIRSPEC Cooling Panel [NIRSPEC-Cool].
-
Remove the CCR line protection hoops. Picture
-
Disconnect the boom tension switch tether rope and cable. Picture
-
Mount and connect TOSS 50
vacuum pump and vacuum gauge to NIRSPEC, using the platform on the
handler frame below NIRSPEC's vacuum valve. Provide 40 foot long AC power
extension cord to pump from Nas Deck connector panel commercial AC outlet.
Start pumping the line only, leaving vacuum valve closed. Ensure vacuum
line is pumped down below 1x10-4 Torr before proceeding.
-
Unplug the NIRSPEC vacuum gauge from NIRSPEC commercial
power and plug it into the extension cord or power strip on the pump.
Verify that the vacuum is being read. Monitor the vacuum throughout
the transfer, and record vacuum readings every 1-2 minutes while CCRs
are off.
-
Obtain a stopwatch.
COMPUTER ROOM
-
Verify that the telescope does not have any handler interlock faults by
switching the telescope drive power on at the MCP. If the power comes on,
there are no Azimuth interlock faults present.
-
Switch the telescope drive power off.
MECHANICAL ROOM
WARNING:
Be expeditious: You have approx. 10 minutes on
the stopwatch to get the NIRSPEC CCR lines hooked up in the AO enclosure
and the compressor started. Designate one person responsible for calling
out stopwatch readings and recording the dewar vacuum every minute. If
the vacuum inside the NIRSPEC dewar exceeds 1x10-3 Torr, check that the
vacuum pump is on and that the vacuum line is pumped down, then open the
vacuum valve on NIRSPEC and begin pumping the dewar.
-
Shut off the NIRSPEC CCR compressor.
-
Immediately start the stopwatch.
RIGHT NASMYTH DECK (in parallel with following tasks in the Computer Room)
-
Disconnect supply and return CCR lines on boom from the NIRSPEC instrument
panel using the tools in the CCR tool kit obtained from the NIRSPEC shelf
in the instrument lab.
-
Disconnect CCR power cable PS1 on boom from NCP,
JS1.
-
Stow boom cables then rotate NIRSPEC boom clockwise to stow position and
slide the dead bolt in place to lock. Picture
COMPUTER ROOM (in parallel with the above tasks on the Right Nasmyth Deck)
-
Disconnect CCR Power Cable PS1 from NIRSPEC Instrument
Interconnect Panel [IIP-NP Right Nasmyth Platform] JS1.
-
Connect CCR Power Cable PS1 to NIRSPEC Instrument
Interconnect Panel [IIP-NP Left Nasmyth Platform] JS1.
TELESCOPE
WARNING: There is a telescope interlock on the
cable boom. If the boom is not properly stowed, the telescope will not
move at its full speed.
-
Rotate telescope in azimuth to 10.8 degrees (which is a move to the
"left" in telescope sense) and roughly align RT4 rails to AO door.
-
Open AO door and fine align RT4 rails to AO rails.
AO ENCLOSURE
-
Drive NIRSPEC handler into the AO enclosure far enough for the
steel-braided CCR hoses to reach their connection points on the
NIRSPEC interconnect panel.
-
Connect the CCR supply and return lines to the NIRSPEC connect panel [NCP].
You will need the CCR tool kit located in the NIRSPEC shelf in the instrument
lab.
-
Connect CCR Power Cable AOCCRAC to JS1 on the NIRSPEC connect panel [NCP].
-
Check vacuum reading on NIRSPEC and note whether
pressure is less than 1x10-3 Torr. If not, then verify vacuum pump is on
and vacuum line is pumped down below 1x10-4 Torr, then open vacuum valve
on NIRSPEC and begin pumping dewar until 1.0e-3 Torr.
MECHANICAL ROOM
-
Turn on the NIRSPEC CCR compressor. Verify the gauge
pressure on the compressor is 128 PSI.
-
Stop stopwatch and note the time and dewar vacuum
for entry in the SIAS.
AO ENCLOSURE
- Note that once the CCR compressor is turned on,
the CCR pressure gauges on the instrument should read 315 to 325 PSI on
the supply line and 125 to 140 PSI on the return line. Oscillation in the
pressure is normal as long as they fall within the range
specified.
-
NOTE ALSO that the CCR cold heads should start
making their familiar "cricket on steroids" sounds immediately upon
restart of the CCR compressors. If they do not, then the CCR cold
head power has not been properly reconnected. Check that the
connections at each end have been properly made:
- NIRSPEC: Check CCR Power Cable AOCCRAC to JS1 on the NIRSPEC
connect panel [NCP].
- Computer Room: Check Power Cable PS1 to NIRSPEC Instrument
Interconnect Panel [IIP-NP Left Nasmyth Platform.
-
Monitor the NIRSPEC dewar vacuum. When it drops below
1x10-3 TORR, close the dewar vacuum valve if necessary, and disconnect
the vacuum pump. Put a blank off on the NIRSPEC inlet valve.
-
Unplug the vacuum gauge from the extension cord/power
strip and plug it into s NIRSPEC commercial AC power outlet.
-
Drive NIRSPEC handler until it stops automatically over the Kinematic mounting
points.
- WARNING: Take care that the underside of the
NIRSPEC handler does not catch on any of the floor insulation
panels.
- DOUBLE WARNING: It has happened at least
once that the position switch fails to stop the NIRSPEC handler
directly over the kinematic mounting points, allowing it to go too far
forward. DANGER, WILL ROBINSON, DANGER! Have someone watch the front
of the NIRSPEC calibration unit as the handler advances. The handler
should stop with the calibration unit 2-3 millimeters before any
contact with the AO bench shroud frame. IF THE CALIBRATION UNIT MAKES
ANY CONTACT WITH ANY PART OF THE AO BENCH, STOP IMMEDIATELY. DO NOT
JACK NIRSPEC DOWN WITHOUT FIRST BACKING UP AT LEAST 1 FOOT AND TRYING
THE FORWARD MOVE AGAIN!
-
Jack NIRSPEC down on to the kinematic mounts by pressing and holding the
"down" button on the side of the handler. Monitor the lift pads on the
rear of the module handler. They should lift slightly off of the handler
frame, then stop automatically. If they fail to stop, release the "down"
button. If the screws drive too far it could damage the lift mechanism.
Additionally, monitor the front screw as it drives down into the kinematic
mount. Ensure that it is seating properly in the hole. Picture
-
Install CCR protection hoops.
-
Remove instrument interface access hole floor panel. Picture
-
Connect NIRSPEC AC Power cable, AOAC to jack on NIRSPEC power/glycol
interconnect panel.
-
Locate and connect the instrument cooling supply and return lines to the
NIRSPEC glycol panel.
-
Open the instrument cooling supply valve, located under the floor opening
on the telescope side of the hole, then open the instrument cooling return
valve show. Picture
-
Connect Fiber patch cables AOFO1 to NICP-J2, and AOFO2 to NICP-J1.
-
Replace the flashing that covers the access hole.
-
Replace the instrument interface access hole floor panel. NOTE:
Ensure all hoses and cables pass through the gap between the floor panel
and the left rail channel.
-
Set the toggle switch on the TOP of the NIRSPEC Calibration Unit to
its Full-Power/AO postion. This switch controls the power to the
internal flat field lamp.
-
Open right front insulation panel on NIRSPEC and turn on main power breaker
on Pulizzi unit. Verify that all 8 green lights in the horizontal row
come on sequentially.Picture
-
Disconnect the handler power from the connector at the RT4 stow position
on the Nas Deck and retract the cable into the takeup reel on the
handler cart.
-
Install AO door threshold and shut and secure AO door. NOTE:
Do not install the floor panel between NIRSPEC and the door as it will
interfere with the handler power cord spool.
-
Pull back the black cloth curtain between NIRSPEC and the AO bench.
Be sure that the curtain is completely pulled towards the electronics
vault (Left side of the AO bench as viewed from the back of NIRSPEC),
and that it is completely out from in front of
NIRSPEC. Otherwise, it will obscure all or part of the beam
entering NIRSPEC.
CONTROL/COMPUTER ROOM
-
Disconnect J14 & J15 (transputer optical fibers)
from the NIRSPEC Instrument Interconnect Panel [IIP-NP], J14 &J15.
-
On the NIRSPEC Instrument Interconnect Panel
[IIP-NP], connect the fiber optic lines for the NIRSPEC transputer as
follows: Connect Fiber J14 to IIP-NP J2; and connect Fiber J15 to
IIP-NP J1.
-
Verify that the "link active" light is lit on the SCSI/Fiber Extender
"Black Box" labelled "NIRSPEC TRANSPUTER" (located adjacent to the
NIRSPEC host computer waimea). If the "link active" light is dark,
cycle power to the Black Box. If the "link active" light remains
dark, check that the fibers are correctly connected.
-
Log on to Sun workstation "kua" as user "nspeceng".
-
From the workspace pulldown menu, choose NIRSPEC Control Menu->Start NIRSPEC
Server (daytime).
-
From the workspace pulldown menu, choose NIRSPEC Control Menu->Check
Server Processes. If the NIRSPEC server is running, 3 processes should
appear in an xterm.
-
Type any key to dismiss the process window xterm.
-
From the workspace pulldown menu, choose Login Windows->xgterm
'waimea' (NIRSPEC) to get a terminal window logged in as user
"nirspec" on the NIRSPEC host computer "waimea."
-
In this terminal window, issue the following commands, in order (each one is
executed when you press Enter):
- setenv NIRSPEC_AO (to tell the software NIRSPEC is
behind AO)
- echo $NIRSPEC_AO (to check that the previous command
worked; should return nothing, just a prompt back; if it says
"undefined variable", stop and call an Instrument Scientist for help).
- m fil1init=1 (...and wait about 30 sec, then)
- s fil1stat (...and repeat every 30 sec or so until status returns "OK")
- m fil1step=27100 (...and wait about 30 sec, then)
- s fil1stat (...and repeat every 30 sec or so until status returns "OK")
- z_init_filters
- exit
These commands will (1) initialize Filter Wheel No. 1 which
contains the AO pupil stop; (2) report the status of Filter Wheel
No. 1; (3) move Filter Wheel No. 1 to the last (as of Feb 2001, it changed
from 27085 when new reflective reimaging optics were installed, RDC)
measured position for aligning the AO pupil stop with the AO pupil;
(4) report the status of Filter Wheel No. 1; (5) initialize Filter
Wheel No. 2, and set the software flags to identify that Filter Wheel
No. 1 is in the AO-pupil-stop position; and (6) quit the window.
-
From the workspace pulldown menu, choose exit...
For Instrument Scientist: CONTROL ROOM or ELSEWHERE WITH NETWORK ACCESS
- Log in as user "nirspec" on the NIRSPEC host computer waimea, and
edit the .cshrc file to UNCOMMENT the line:
#setenv NIRSPEC_AO
- Log in as user "nirspec" on the Remote Ops 2 Observer's machine
hanauma, and edit the .cshrc file to UNCOMMENT the line:
#setenv NIRSPEC_AO
last modified: D. Sprayberry, 2 May 2001