![]() |
Troubleshooting An Overheating PXL Camera |
|
All the error messages described in this page will be seen by the OA, not by the observer. Also, all the actions described here must be taken by the OA. These actions will probably require an interruption of observing to get the PXL camera working again.
The original source of this information is Nightlog Ticket K2-04864.
The OA sees a number of error messages appearing on his or her console, generated by the waimanu cam_server program:
waimanu cam_server [pid]: pxl current temp too warm [T]
where [T] is number. NOTE that the number given as [T] must be divided by 100 to get the actual temperature of the camera in degrees C.
NOTE also that the normal temperature set point for the NIRSPEC PXL camera is -25C.
When this problem occurs, the OA should take the following actions:
This type of problem has been reported twice. These problems are most likely due to some stuck bit or serial communications glitch between the controller and the camera. Resetting the controller as described has cleared the problem both times.
Note that this type of problem, and these error messages from the cam_server, are not due to a glycol flow problem. However, it is still a good idea to check the glycol flow, just to eliminate that as a possible additional problem.