Possibly due to Pupil Mask Rotator (PMR) not operating
Solution
Check the NIRC2 status gui. If you are using one of the Hex
patterns, and either tracking is false, or, the angle is fixed
at 0.0, then likely the problem is due to misalignment of the
hex mask. If tracking is false, start tracking with "pmrrot on".
If tracking is true, but the angle is stuck at 0.0, restart
the daemon with "nirc2 start pmr"
Chevron noise pattern as seen in left hand side of
the two images shown below (displayed with ds9 using histeq).
Problem
Possibly caused by the filter animatics motor staying in an
energized state because it did not quite reach its destination.
This seems to happen more frequently when selecting narrow band
filters.
Solution
Select brgamma ("filt brgamma"), select Kp ("filt Kp"), and take
another dark. The result should look more like the nominal noise
pattern that is shown on the right hand side of the above ds9
figure.
Vignetting over 1/3 of the array is apparent
in wide camera images. Much lower vignetting
may also be present in the narrow camera,
but the narrow camera effects are difficult
to detect.
Above: Wide camera flat image with vignetting present, and a diagonal cut across the image from upper left to lower right.
Cannot show or modify alad keywords and the exposure meter
is showing negative numbers or there are negative numbers
in a waikoko xterm where the exposure command was issued.
Problem
There is something wrong with the alad server. Probably due to <3 alad servers running.
Solution
Restart the alad server (as nirc2eng on waikoko)
nirc2 stop alad
Wait 5 seconds
nirc2 start alad
Check servers using a "ct" or "testAll"
If this fails because motor and IO daemons not running,
restart all three (io,motor,alad) with "nirc2 start deamons";
but see below for possible need to re-home and reset cold head speeds.
If the transputer check fails, see
GH instructions on OA page
for power sequencing.
Another useful document for any proplems with terminal servers,
etc is the
NIRC2 gotchas document. Also consult
Alad Troubleshooting Tips & Tricks
Cannot show or modify motor keywords ("Unable to read NIRC2 THEAD2LO" email alarm)
Problem
Likely due to power failure.
Solution
Restart the daemons and, if necessary reset cold speed and re-home motors.
nirc2 start motors (starts both IO and motor daemons)
Check motors with showmot (false values under home imply need to re-home)
If motors are homed skip down to cold steps.
Rehome the motors as follows:
For all motors except camera and slit slides(i.e., FWI,FWO,GRS,SHR,PMR,PMS,PSI,PSO):
mm xxxRESET=1
mm xxxINIT=1
mm xxxHOME=1
For sls/slm/cam, use the poshome command after the reset
and init. Check homelog.070220
to be sure behavior is consistent. Try not to home SLM/SLS
to negative limits (see RC for details). Update the log