If you take dome flats in the morning after your observing night, be sure to use the commands endnight to put NIRC2 into safe mode and aohatch close to protect the adaptive optics bench when you are finished.

You can obtain at least broad-band flat fields by using the dome flat lamps designed for the optical instruments. Ask the OA to point to a good flat field spot on the dome (usually at an elevation of 45°) and turn on the spectroscopic lamps. These are the bright lamps. Below is a table of exposure time estimates that should produce 8000 DN or somewhat less in each broadband filter and camera combination. Numbers in parantheses indicate that the exposure time is to short to reach with the full frame. Either a subarray can be used, or you can try using a flat field from the narrow camera to calibrate data taken with the wide or medium camera. Note that this technique has NOT been tested; it is only a suggestion.

Note also that there are two flat field lamps. If you see only 1/2 of the expected flux, there may be a burned out lamp. Ask the OA to check by looking for uneven illumination on the dome.

Filter
Camera
wide
narrow
J
1.2
20
H
1.2
20
K
6
150
Kp
3
45
Ks
4.5
70
Jcont
?
?
Pa Beta
10
160
Hcont
10
160
CH4_long
5
80
Fe II
15
240
CH4_short + PK50_1.5
-
30
He I B
25
400
Br gamma
90
160
H_2 1-0
100
1600
H_2 2-1
300
4800
Kcont
450
7200
CO
300
4800
Br gamma2
90
1600
H2O
1
16
PAH
1
16

Estimated exposure times in seconds for NIRC-2 dome flats.

DCS keywords for lamp control = flimagin and flspectr
(e.g., to turn both off "modify -s dcs flimagin=0 flspectr=0")

 

 

Al Conrad