From goodrich Thu Mar 8 18:10:21 2001 To: goodrich, aconrad Subject: just to get it written down... Major outstanding issues: subc Failures were reported, but things have looked good from Waimea? snapi Needs to get checked. corona I haven't added the positional argument. Do we want to? We could require observers to center a target on the spot by moving the field, rather than the coronagraphic spot. If the spot moves too far, you'll get another spot (or something) coming in to the edge of the field, possibly. various Things with valid lists of positions, like "filt" and "fgr" have not all had those lists added to the help text. Some have no valid positions yet (like fgr and sgr). preslit Still potential problems with determining whether there is a slit in the beam or not. I have fixed the "image" command, however, which did not reset "slitname", and this will take care of the specific instance which Eiichi documented. sgr Works OK, but documentation needs to be changed. slit Still apparently a problem in getting the Status Window to update. grand Does not produce random sequences. Al needs to find a better random number generator. lastframe Does not warn against potential overwrite situations. It would need to check the FILEOVWR keyword to see whether it is on or not. If not, then file overwriting will not occur; the frame number will simply be advanced to the next highest number and write a new file, negating the effects of a frame command. (Note that users are not supposed to use "lastframe", but rather "frame".) Bob