Detector Summary |
|
This page provides general information on the LWS detector.
| Manufacturer | Boeing |
| Format | 128×128 pixels |
| Composition | Si:As moderate flux array |
| Pixel size | 75 µm |
| Pixel scale | 0.0847 arcsec/pixel |
| Dark current | 5×105 e-/sec/px |
| Read noise | 1500 e- |
| Inverse Gain | 2222 e-/ADU |
| Full Well Capacity | 1.1 × 107 e-/pix |
| ADC range | 12 bit |
| Bias level | 100 ADU |
| DQE | 0.7 |
| Non-linearity | <0.01% over entire ADC range |