The proper focus of each camera varies with filter and observing mode (e.g., direct imaging and slitmask spectroscopy vs. longslit spectroscopy) and, for the red side, with the dichroic in use. Even though the zero point for the focus changes daily, the relative offsets between various focus modes on LRIS are believed to be stable. Hence, if the observer focuses the instrument in the nominal imaging mode then the focus for all other instrument modes can be derived from that measurement.
This document describes the officially-endorsed procedure for daily focusing of LRIS. The procedure involves using a pinhole mask in the LRIS focal plane to obtain a series of images with varying focus values, then analyzing these images to derive the optimal focus for imaging mode. Once the instrument is focused, the observer generates a table showing the corresponding focus values in other modes.
LRIS Utilities > Run Focusloops > DOME lamps, blue & red (DEFAULT)The above is the preferred method. Other methods for focusing the instrument are found in the Run Focusloops pulldown area on the desktop menu. In particular,you could focus the instrument internally by selecting:
LRIS Utilities > Run Focusloops >INTERNAL lamps, blue & red (DEFAULT)This scripts will:
Press <Enter> to exit...Press the Enter key to terminate this window.
LRIS Utilities > Xfocus > Xfocus red
LRIS Utilities > Xfocus > Xfocus blue
|
|
LRIS Utilities > Print Focus TableThe focus table will be printed on the printer in Remote Ops I.
focus_tableor print it by executing either
focus_table | lpr -Plw4
focus_table | lpr -Plw6
This table is only valid if you are starting in a configuration known
to be in good focus. Typicaly you would run xfocus in some
configuration, set to the recommended focus and then run "focus_table".
The current setup is read, tabulated offsets calculated from the
red and blue filters, red and blue focus values, the dichroic, and
whether there is a long slit in or not. Then new values are calculated
for all other combinations.
R E D S I D E F O C U S
Filter Imaging Longslit
+ multislit spectroscopy
+ superlong
==============================================
clear 2287 2246
B 2182 2141
V 2387 2346
R 2227 2186
Rs 2252 2211
I 2233 2192
GG495 2281 2240
OG570 2175 2134
RG850 2279 2238
Dichroic add offset:
mwmwmwmwmwmwmwmwmwmwmw
clear 0
mirror 0
460 -7
500 11
560 16
680 -9
B L U E S I D E F O C U S
Filter Imaging Longslit
+ multislit spectroscopy
+ superlong
==============================================
clear -2408 -2373
U -2383 -2348
B -2428 -2393
G -2408 -2373
V -2407 -2372
NB4170 -2433 -2398
There is no dichroic offset for the blue side