High Resolution Echelle Spectrometer
Technical Page
|
Summary
- Symptom
- When a cross disperser move is requested, the XDANGL keyword
starts changing and never stops; it cycles continuously
within a certain range of values.
- Problem
- The cross disperser motor is confused.
- Solution
-
- From the window manager's pulldown menu, select the
HIRES Control Menu option, then release on the
item labeled Stop Cross-disperser Oscillation.
- Alternatively, type the following command on
lehoula:
stop_cd_drift
- If this does not halt the cross-disperser, contact your
support astronomer.
- Symptom
- During a non-dark, exposure the color-coded box on the expose GUI
stays black indicating a closed shutter.
- Problem
- From what we can tell so far (Sep 2006) the shutter is actually
opening, but somehow the switch is not quite making so that the
software thinks it is closed.
- Solution
-
- No solution needed, unless the exposure level implies that
the shutter really was closed. This is something we have
never seen happen (Sep 2006).