| Deckname |
Length (") |
Width (") |
Resolution (calculated) |
Resolution (measured*) |
Resolution (measured+) |
| B1 |
3.5 |
0.574 |
72,000 |
67,000 |
66,400 |
| B2 |
7.0 |
0.574 |
72,000 |
67,000 |
66,400 |
| B3 |
14.0 |
0.574 |
72,000 |
67,000 |
66,400 |
| B4 |
28.0 |
0.574 |
72,000 |
67,000 |
66,400 |
| B5 |
3.5 |
0.861 |
48,000 |
49,000 |
50,000 |
| C1 |
7.0 |
0.861 |
48,000 |
49,000 |
50,000 |
| C2 |
14.0 |
0.861 |
48,000 |
49,000 |
50,000 |
| C3 |
28.0 |
0.861 |
48,000 |
49,000 |
50,000 |
| C4 |
3.5 |
1.148 |
36,000 |
37,000 |
37,500 |
| C5 |
7.0 |
1.148 |
36,000 |
37,000 |
37,500 |
| D1 |
14.0 |
1.148 |
36,000 |
37,000 |
37,500 |
| D2 |
28.0 |
1.148 |
36,000 |
37,000 |
37,500 |
| D3 |
7.0 |
1.722 |
24,000 |
24,000 |
24,700 |
| D4 |
14.0 |
1.722 |
24,000 |
24,000 |
24,700 |
| D5 |
0.119 |
0.179 |
pinhole |
| E1 |
1.0 |
0.400 |
103,000 |
84,000 |
86,600 |
| E2 |
3.0 |
0.400 |
103,000 |
84,000 |
86,600 |
| E3 |
5.0 |
0.400 |
103,000 |
84,000 |
86,600 |
| E4 |
7.0 |
0.400 |
103,000 |
84,000 |
86,600 |
| E5 |
1.0 |
0.800 |
51,000 |
52,000 |
52,000 |
-
* Using UV cross-disperser. Average of 5 Th/Ar lines near 4100 A.
-