Transcript of a Bad ESI CCD Crate Reboot

This is an example of an ESI crate reboot the failed, resulting in analog power not being supplied to the CCD crate. See the text in red below.


-> reboot
reboot

Press any key to stop auto-boot...
 7 6 5 4 3 2 1 0
auto-booting...

boot device          : ln
processor number     : 0 
host name            : esinstb.ucolick.org
file name            : /vx/vw/vxWorks
inet on ethernet (e) : 192.168.5.5:ffffff00
host inet (h)        : 192.168.5.3
user (u)             : kics
flags (f)            : 0x2 
target name (tn)     : esi5cb.ucolick.org
startup script (s)   : /vx/esi5ce/startup.cmd

Attaching network interface ln0... done.
Attaching network interface lo0... done.
Loading... 476728 + 32904 + 24240
Starting at 0x20000...

Attaching network interface ln0... done.
Attaching network interface lo0... done.
Loading symbol table from esinstb.ucolick.org:/vx/vw/vxWorks.sym ...done
 

 ]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]
 ]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]
 ]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]
      ]]]]]]]]]]]  ]]]]     ]]]]]]]]]]       ]]              ]]]]         (R)
 ]     ]]]]]]]]]  ]]]]]]     ]]]]]]]]       ]]               ]]]]            
 ]]     ]]]]]]]  ]]]]]]]]     ]]]]]] ]     ]]                ]]]]            
 ]]]     ]]]]] ]    ]]]  ]     ]]]] ]]]   ]]]]]]]]]  ]]]] ]] ]]]]  ]]   ]]]]]
 ]]]]     ]]]  ]]    ]  ]]]     ]] ]]]]] ]]]]]]   ]] ]]]]]]] ]]]] ]]   ]]]]  
 ]]]]]     ]  ]]]]     ]]]]]      ]]]]]]]] ]]]]   ]] ]]]]    ]]]]]]]    ]]]] 
 ]]]]]]      ]]]]]     ]]]]]]    ]  ]]]]]  ]]]]   ]] ]]]]    ]]]]]]]]    ]]]]
 ]]]]]]]    ]]]]]  ]    ]]]]]]  ]    ]]]   ]]]]   ]] ]]]]    ]]]] ]]]]    ]]]]
 ]]]]]]]]  ]]]]]  ]]]    ]]]]]]]      ]     ]]]]]]]  ]]]]    ]]]]  ]]]] ]]]]]
 ]]]]]]]]]]]]]]]]]]]]]]]]]]]]]]
 ]]]]]]]]]]]]]]]]]]]]]]]]]]]]]       Development System
 ]]]]]]]]]]]]]]]]]]]]]]]]]]]]
 ]]]]]]]]]]]]]]]]]]]]]]]]]]]       VxWorks version 5.2
 ]]]]]]]]]]]]]]]]]]]]]]]]]]       KERNEL: WIND version 2.4
 ]]]]]]]]]]]]]]]]]]]]]]]]]       Copyright Wind River Systems, Inc., 1984-1995

                               CPU: FORCE CPU-5CE.  Processor #0.
                              Memory Size: 0x700000.  BSP version 1.0.

Executing startup script /vx/esi5ce/startup.cmd ...
loginUserAdd "kics","S9deyRSb9S"
value = 0 = 0x0
nfsMount "esinstb.ucolick.org","/vx","/vx"
value = 0 = 0x0
nfsAuthUnixPrompt
machine name:   esinstb.ucolick.org user ID:        2001 group ID:       100 num of groups:  0 value = 0 = 0x0
iam "kics"
value = 0 = 0x0
ifBroadcastSet "ln0", "192.168.5.255"
value = 0 = 0x0
hostAdd "esupvb.ucolick.org","192.168.5.4"
value = 0 = 0x0
hostShow
hostname         inet address       aliases
--------         ------------       -------
esi5cb.ucolick.org 192.168.5.5       
localhost        127.0.0.1         
esinstb.ucolick.org 192.168.5.3       
esupvb.ucolick.org 192.168.5.4       
value = 0 = 0x0
nfsDevShow
device name          file system                                       
-----------          -----------                                       
/vx                  esinstb.ucolick.org:/vx                           
value = 0 = 0x0
cd "/vx/esi5ce"
value = 0 = 0x0
< logo
cd "/vx/kroot/kss/esi/vme/force5ce/vmeinf"
value = 0 = 0x0
vmemsgxchng:  DSP reply timeout.  Check power/hardware.
value = -1007 = 0xfffffc11 = sra.o_bss + 0xff0236b9
taskDelay(15)
E1:Time=2017:E1:Time=2017:value = 0 = 0x0

cd "/vx/kroot/kss/esi/vme/force5ce/vmeinf/lick"
value = 0 = 0x0
ld < tim_download.o
value = 15975456 = 0xf3c420 = _sline + 0x2c0
ld < tim_verify.o
value = 16247176 = 0xf7e988 = _sline + 0x48
ld < vme_download.o
value = 16245600 = 0xf7e360 = service_request.o_bss + 0x190
ld < vme_verify.o
value = 16243952 = 0xf7dcf0 = tb_com.o_bss + 0x408
ld < util_download.o
value = 16243392 = 0xf7dac0 = tb_com.o_bss + 0x1d8
ld < util_verify.o
value = 14852968 = 0xe2a368 = util_verify.o_bss + 0x690
ld < startup.o
value = 14845816 = 0xe28778 = startup.o_bss + 0x8

cd "/vx/kroot/kss/esi/vme/force5ce/ldserv"
value = 0 = 0x0
ldvmemsgxchng:  DSP reply timeout.  Check power/hardware.
error -1007 reading 804119
verify error: at address 804119  expected: 000aad0a  got:  00000000
E1:Time=3046:TIMING board verification status= 2, should be 0
Will verify utility eprom against /vx/esi5ce/dsp/utility/util.s
Expecting 1 verification error: at 800081
verify error: at address 800081  expected: 000004  got: 052d79
EOF
UTILITY board verification status= 1, should be 1
EEPROM verification failure; power-up inhibited
value = -3 = 0xfffffffd = sra.o_bss + 0xff023aa5

taskSpawn("broad_mon", 100, 0x19, 10000, broadcast_mon)
value = 14834256 = 0xe25a50
msg size = 2 words.  reply size = 2 words
msg = 0X302 0X48485452
reply = 0X30002 0X444f4e
dewar heater DAC value reset to 1638
Digital input has changed. Old value=0  New value=fef3
Digital input (mask ID 1) has changed. Old value=ffffffff  New value=1
Digital input (mask ID 1e) has changed. Old value=ffffffff  New value=9

digital input broadcast:  [0]= 2  [1]= 1  [2]= 1  [3]= 30  [4]= 9 
Digital output has changed. Old value=0  New value=4096
Digital output (mask ID 1024) has changed. Old value=-1  New value=0
Digital output (mask ID 32768) has changed. Old value=-1  New value=0
Digital output (mask ID 16384) has changed. Old value=-1  New value=0
Digital output (mask ID 512) has changed. Old value=-1  New value=0
Digital output (mask ID 256) has changed. Old value=-1  New value=0

Done executing startup script /vx/esi5ce/startup.cmd

Last modified: Tue Jan 17 20:49:15 HST 2006