NIRC2 Instrument and Data Format*

1. Introduction

NIRC2 is a near-infrared instrument designed to take full advantage of the adaptive optics on the Keck II telescope of The W. M. Keck Observatory on Mauna Kea, Hawaii.The Co-PIs are Keith Matthews and Tom Soifer, both of Caltech. The instrument was built by Keith Matthews and engineer Sean Lin of Caltech, with help from James Larkin, Ian McLean, and others at UCLA (detector electronics and related software), and Al Conrad, Bob Goodrich, and Allan Honey at Keck Observatory (software). Support in Waimea was provided by Jim Bell, Randy Campbell, and Drew Medeiros. It was delivered to Hawaii in the spring of 2001, seeing first light in the summer of 2001.

NIRC2 is positioned behind the AO bench on the Left Nasmyth Platform of Keck II. The instrument operates from 1 to 5 µm, providing three selectable cameras to cover image sizes between 10"x10" and 40"x40". Two filter wheels with 18 positions each provide a variety of filters and/or grisms, while a focal plane mechanism provides slits and occulting spots for coronography. A dedicated slide carries larger grisms for spectroscopy. Six selectable pupil masks are available to reduce background noise sources; four of these rotate in concert with the telescope pupil and one is specific to spectroscopy. The detector is a 1024x1024 Aladdin-3 InSb array with four-quadrant readout into 32 channels.

2. Optical path

The light provided by the AO system enters the NIRC2 dewar through a CaF2 window. It then passes through two preslit blades which provide the first level of baffling. These movable blades change position depending on which camera is in use and whether or not a slit is in the beam. After the preslits the light is collimated. The beam comes to a focus where there are two slide mechanisms. The slit slide carries several slits of varying widths. A mask slide beneath it selects which slit provides light to the rest of the instrument. The slit slide also contains a set of coronographic spots. A third use of the slit and mask slides is to provide a focal plane baffle for imaging through the three cameras.

Beneath the slit is a wheel which provides a selection of pupil masks, including one for spectroscopy, and two rotating masks designed to block light from the spiders as well as the secondary obstruction. Next comes two filter wheels with 18 positions each. Below the filter wheels is a relatively slow shutter, which can be used to close off external light and may also be used to stop an exposure if, for example, the AO system loses lock on a target. The beam next passes through a grism slide, with three slots plus an open position.

Last in the optical path is one of three selectable cameras. These have nominal scales of 0.01, 0.02, and 0.04 arcsec/pixel, matching the expected diffraction performance of the telescope and AO system at different wavelengths (J, K, and M bands, roughly). The cameras focus onto the InSb detector, described in more detail below.

3. Mechanical description

All moving parts in NIRC2 are inside the dewar. The filter wheels and pupil mask are on rotating wheels, while other mechanisms (preslits, slits, coronagraphic spots, grisms, the shutter, and the cameras) are on slide mechanisms.

Cooling is provided by a closed-cycle system similar to that installed on NIRSPEC. Two cold heads provide three stages of cooling, controllable via software. Temperature sensors are provided on the detector head, optical bench, cold shield, and other places in the dewar.

4. Detector and Electronics

The detector is a 1024x1024 Aladdin-3 array with 27 µm pixels. The detector QE is ~ 80% at 1.7 microns.

The electronics provide three readout modes: single sample, CDS (correlated double sampling), and MCDS (multiple correlated double sampling) modes. The pixel sampling rate is nominally 250 kHz, but can be changed up to ~350kHz. Each quadrant is read out in eight channels (32 channels total). The gain of the detector is 4 electrons/DN.

The read noise in a CDS readout is 38 electrons. For MCDS mode with up to 64 reads, readout noise drops as the square root of the number of reads. Beyond that the noise continues to drop, although not as rapidly as the square root. It continues to drop to 512 reads, beyond which it has not been tested. Section 6.2 shows the read noise as a function of number of MCDS reads.

The detector is linear to about 10,000 DN. This is the result of a compromise between linearity and dark current.

5. Observation modes

NIRC2 has three basic modes.

5.1 Imaging

Narrow- and broad-band filters are provided. The filter complement is given in the filter page, together with sensitivity and background values to aid in signal-to-noise estimates.

Three different cameras provide scales of 0.01, 0.02, and 0.04 arcsec/pixel. These provide good matches to the Keck diffraction size at J, K, and M bands.

The main concerns in imaging mode are image scale or field of view, filters, exposure times, and dither patterns. The "camera" command is used to select one of the three cameras in NIRC2.

Table 1: Camera Options

Camera
Name(s)
Scale
(arcsec/pixel)
Field of view
(arcsec x arcsec)
Band in which
diffraction size
matches 2 pixels
wide
cam04
0.04
0.04
40x40
M
medium
cam02
0.02
0.02
20x20
K
narrow
cam01
0.01
0.01
10x10
J

Note that the camera command will take different names; it might be easier to remember the pixel scale rather than the relative field size, for example.

Also note that the medium-field camera has significant field curvature, hence image quality across the entire field suffers. It is possible to achieve high image quality for point sources, however.

For imaging, when a new camera is selected, the appropriate masks in the slit and slitmask slides are automatically put into place.

5.2 Coronagraphy

Occulting spots can be placed in the focal plane using the slit slide. The spots are opaque areas deposited on a quartz substrate. The spot transmissions are not zero, but have been measured to be 0.12% in both J and H, and 0.22% in K. (These numbers are relatively uncertain). A variety of spot sizes are available. The following lists their names, with the number after "corona" being the spot size in milliarcseconds:

corona100
corona150
corona200
corona300
corona400
corona600
corona800
corona1000
corona1500
corona2000

5.3 Spectroscopy

Grisms may be installed either in the filter wheel or in a dedicated grism slide. Currently there are two grisms, both in the grism slide. The tables below list various useful parameters for NIRC-2 spectroscopy. The columns are:

  1. Camera
  2. Filter
  3. Order
  4. Central wavelength of the NIRC-2 filter (microns)
  5. Dispersion (Angstroms/pixel; remember that 1 Angstrom is 0.0001 microns)
  6. Wavelength range of the filter (microns)
  7. The number of pixels needed to cover the entire wavelength range. Bold red letters indicate that the entire filter passband will fit on the detector.
  8. The resolving power per pixel. If you use a three-pixel wide slit, e.g., your resolving power would be 1/3 that listed.

Table 2: LOWRES grism

Camera
Filter
Order
Central
wavelength
(microns)
Dispersion
(Ang./pix)
Wavelength
range of
filter
(microns)
# pixels
to cover
range
Resolving
power
per pixel
Narrow
J
6
1.248
1.092
0.163
1490
11,430
J
5
1.248
1.308
0.163
1250
9,540
H
4
1.633
1.628
0.300
1840
10,030
K
3
2.196
2.144
0.336
1570
10,240
Kp
3
2.124
2.144
0.336
1640
9,910
Medium
J
6
1.248
2.190
0.163
740
5,700
J
5
1.248
2.616
0.163
620
4,770
H
4
1.633
3.255
0.300
920
5,020
K
3
2.196
4.113
0.336
820
5,340
Kp
3
2.124
4.113
0.336
850
5,160
Wide
J
6
1.248
4.373
0.163
370
2,850
J
5
1.248
5.236
0.163
310
2,380
H
4
1.633
6.519
0.300
460
2,500
K
3
2.196
8.226
0.336
410
2,670
Kp
3
2.124
8.226
0.336
430
2,580

Table 3: MEDRES grism

Camera
Filter
Order
Central
wavelength
(microns)
Dispersion
(Ang./pix)
Wavelength
range of
filter
(microns)
# pixels
to cover
range
Resolving
power
per pixel
Narrow
J
6
1.248
0.644
0.163
2530
19,380
J
5
1.248
0.704
0.163
2320
17,730
H
4
1.633
0.929
0.300
3230
17,580
K
3
2.196
1.267
0.336
2650
17,330
Kp
3
2.124
1.267
0.336
2770
16,760
Medium
J
6
1.248
1.288
0.163
1270
9,690
J
5
1.248
1.407
0.163
1160
8,870
H
4
1.633
1.858
0.300
1610
8,790
K
3
2.196
2.534
0.336
1330
8,670
Kp
3
2.124
2.534
0.336
1390
8,380
Wide
J
6
1.248
2.576
0.163
630
4,840
J
5
1.248
2.797
0.163
580
4,460
H
4
1.633
3.420
0.300
880
4,770
K
3
2.196
5.078
0.336
660
4,320
Kp
3
2.124
5.078
0.336
690
4,180

 

Slits are also available in a variety of widths. A list is given below, where the number after "slit" gives the width in milliarcseconds.

slit10
slit20
slit30
slit40

slit60
slit80
slit120
slit160

Note that since the slits and coronagraphic spots are on the same slide, it is not possible to do coronagraphic spectroscopy.

6. Observing with NIRC2

6.1 Sensitivities/backgrounds

Below is a table listing all filters available with NIRC2, including zeropoint and sky estimates (when measured), and the maximum exposure time expected in the narrow-field camera (before the sky saturates). See the NIRC2 webpage for the latest suite of filters.

Table 4: Filters, sensitivities, and background rates
(Clicking on a filter name will bring up a plot of that filter's transmission curve.)

Filter
Central
wavelength
(µm)

Bandpass
width
(µm)

Cut-On Wavelength (µm)

Cut-Off Wavelength (µm)

Photometric
zeropoint
(Note: This is for Strehl = 1!!)

Sky
(mag./
sq. arcsec)
T(max)
(sec)
Broad Band
1.0311
0.0481
1.007
1.0551
     
1.0180
0.0996
0.9682
1.0678
     
1.248
0.163
1.166
1.330
25.35
14.9
10,000
1.633
0.296
1.485
1.781
25.44
13.6
2750
2.196
0.336
2.028
2.364
24.63
12.6
630
2.146
0.311
1.991
2.302
24.53
12.2
630
2.124
0.351
1.948
2.299
24.74
12.2
630
3.5197
1.3216
2.8589
4.1805
     
3.776
0.700
3.426
4.126
23.2
2.91
0.27
4.670
0.241
4.549
4.790
21.2
-0.12
0.14
Narrow Band
1.0847
0.0182
1.0756
1.0938
     
1.096
0.016
1.088
1.104
     
1.2132
0.0198
1.2033
1.2231
1.2903
0.0193
1.2807
1.3000
 1.5804
0.0232
1.5688
1.5920
 
 
 
1.5923
0.1257
1.5295
1.6552
1.6455
0.0256
1.6327
1.6583
 
 
 
1.6809
0.1368
1.6125
1.7493
2.0563
0.0326
2.0400
2.0726
2.1686
0.0326
2.1523
2.1849
 
 
 
2.1281
 0.0342
2.1112
2.1452
 
 
 
H2 v=2-1
2.2622
0.0388
2.2428
2.2816
     
2.2706
0.0296
2.2558
2.2854
 
 
 
2.2891
0.0267
2.2757
2.3024
3.0629
0.1549
2.9855
3.1404
3.2904
0.0555
3.2627
3.3182
3.987
0.069
3.952
4.021
     
4.052
0.068
4.018
4.086
     

 

 

The NIRC2 Zero point values were updated after photometric tests in April 2004.

This table shows the zero-point, Z, and sky brightness, S, through various NIRC2 filters, combined with the narrow camera and the circumscribed pupil (the "open" position of the pupil wheel).


Z is the magnitude of the star + 2.5 log10 (counts/sec) - 2.5 log10(Strehl). Note that the Strehl ratio will vary according to atmospheic conditions (e.g. seeing), the brightness of the AO lock star, and the distance away from the lock star.

The gain, G, is 4.0 electrons/count.

S is the sky brightness in magnitudes per square arcsec. The sky background in counts per pixel is given by:

counts/sec/pixel = (arcsec/pixel)^2 * 10^[0.4*(Z-S)]

where (arcsec/pixel) is the pixel scale, e.g. 0.040 for the wide camera.

T(max) is the maximum exposure time without saturating the background in the narrow camera.

For different pupil masks:

Inscribed circle, subtract 0.4 from the zeropoints above, subtract 0.5 from the sky backgrounds above.
The other pupil masks have not been measured, but they should lie between these two extremes.

For different cameras:

T(max) should be divided by 4 for the medium camera, and 16 for the wide camera.

Notes:

a) The thermal bands have not been corrected for air mass. This is a future measurement.

b) The grisms have not been measured. A rough efficiency (not including slit losses) is 50% for the grism throughputs (slightly better at H and K).

 

6.2 Noise sources

Noise sources include noise in the electronics, statistical noise from the background, and statistical noise from the target itself. Which of these noise sources dominates depends on your observing mode. Bright targets will often be dominated by their own statistical noise, while faint targets may be dominated by background noise. In the thermal infrared (beyond ~ 2.5 µm) almost all observations are dominated by background.

Read noise depends on the readout mode of the detector. In CDS mode it is 38 electrons/pixel. In MCDS mode it drops roughly as the square root of the number of reads until around 64 reads, then drops more slowly until at least the 512-read level. A plot of the read noise vs. number of MCDS samples is shown below:

Figure 1. Read noise vs. number of MCDS samples.

Statistical noise from the target and background are simply the square root of the number of electrons recorded from the source.

Dark current is also a function of the number of MCDS samples. This is because every time the detector is read the four on-chip amplifiers glow slightly. This glow shows up as an elevated background and rises in the background in the four corners of the image. The figure below shows the dark current (in DN/sec) as a function of the number of MCDS reads for a 200 sec. exposure. The drop in effective dark currrent above 300 samples is not understood.

Fig. 2. Dark current vs. number of MCDS samples (200 sec exposure).

6.3. Instrument Control

Control of NIRC2 is done through a command line interface (CLI). The instrument computer is named "waikoko" and is on the summit. Almost all instrument-related software runs remotely on waikoko, and almost all instrument-related commands must be typed into a waikoko xterm.

There are numerous commands for NIRC2, but many of them are considered engineering commands, others are low level commands used by higher level commands, and still others are expected to be rarely, if ever, needed by the observer.

The instrument status is displayed in a window showing the most useful NIRC2 parameters:

Figure 3. The NIRC2 status display.

The image display GUI is known as "QuickLook," similar to one for NIRSPEC. In general it will automatically display images as they are taken, with one important caveat. During rapid sequences of images some may be "skipped," as the software tries to keep in step with the data flow.

Figure 4. The QuickLook image display and analysis tool.

The QuickLook GUI itself has a number of important observing features, including allowing observers to make small telescope moves, measuring the Strehl ratio on images, and displaying pair-subtracted images.

7. NIRC2 Image Format

All NIRC2 data files are two-dimensional images that are stored in standard simple FITS format consisting of a single primary Header/Data Unit (HDU) without any extensions. The image pixels are in 32-bit integers and are 1024 x 1024 long for a full-frame, un-windowed image. For a description of keywords in the NIRC2 header unit, visit http://www2.keck.hawaii.edu/koa/public/keywords/nirc2Keywords.php. In spectroscopy mode, the wavelength dispersion direction of the spectral image is along the X axis (NAXIS1). A full-frame raw image contains 4216320 bytes.

Some examples of typical NIRC2 images and spectra are shown below.

Figure 5 : Example of a NIRC2 direct image

 

Figure 6: Examples of NIRC2 two-dimensional spectral image of a star (left) and an arc lamp (right) showing bright emission lines. Wavelength increases from left to right.

8. General Specifications

The following table shows the current general specifications for NIRC2. For the lastest information, check the NIRC2 webpage.

Table 5. General Specifications

Input f/ ratio f/15
Wavelength range 0.9-5.3 microns
Field of view 10x10 arcsec
20x20 arcsec
40x40 arcsec
Pixel scale

0.01 arcsec/pixel
0.02 arcsec/pixel
0.04 arcsec/pixel

Filters Z, Y, J, H, K, Ks, Kp, Lw, Lp, Ms, HeIA, HeIB, Pa_gamma, Jcont, Pa_beta, Hcont, H2, CH4S, CH4L, Fe II, Kcont, CO, H2O, PAH, Br_alpha, Br_alpha_cont
Pupil mask two circular fixed,
four hexagonal, rotating
Slits 10, 20, 30, 40, 60, 80, 120, and 160 milliarcsec
Coronagraphic spots circular: 100, 150, 200, 300, 400, 600, 800, 1000, 1500, 2000 milliarcsec
Grisms medres and lowres grisms (see tables 2 & 3), pupil imaging lens, and clear (imaging mode).
Minimum frame time 0.182 sec, full frame
Detector 1024x1024 InSb Aladdin-3 array, 27 micron pixels
32 simultaneous outputs, grouped in 8 pixels across a row in each of 4 quadrants
gain ~ 5 electrons/DN
subarray capability
35 K operating temperature
Read noise 38 electrons/read
11.5 DN/sqrt(N) for N samples in MCDS mode, up to N = 64, dropping more slowly beyond that until at least N = 512. (Note that N = 2 corresponds to two reads, or CDS mode.)
Dark current < 0.1 electrons/pixel/sec
QE 80% at 1.7 microns
Well depth 100,000 electrons (at 1% linearity?)

 

* From the NIRC2 User's Manual with updates by the KOA team